ESR 60 min T.Corr
Temperature correction is used. This is the 60 minutes value
corrected to 18°C.
Date / time
Date and time of the measurement of the result.
ESR time (min.)
Actual duration of the ESR.
Dilution %
The calculated dilution rate after aspiration of the sample.
Temperature (°C)
Room temperature at the time of the measurement of the sample.
Pipet number
Pipette in which the sample was measured.
Error code
Shows any ESR error or warning code (e.g. "Too many borders
found").
A complete overview of measurement data is shown with activating the option "Display raw data".
6.3.3 - ESR Statistics screens
A statistical graph is produced over a selected period. Make a selection of the following graphs;
•
Daily mean (mm)
Use this to check variations in the daily mean ESR.
•
Hazy (%)
Increasing hazy aspects are an indication for contamination of the instrument, see
•
Dilution (%)
Increasing dilution errors indicate the need for maintenance of the diluter system.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
49
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...