11
Waste disposal
The waste container has a level sensor and as soon as the level sensor generates a waste error, the waste
container must be emptied. The waste must be treated as potentially infectious (biohazardous) material and
disposed of according to local regulations. Preferably, discard the complete waste container and replace it
with a cleaned one. Clear the error.
If the waste line is to be connected to a centralised waste collection system, the following requirements
must be met:
1. Waste tube must not exceed 5 meters or 18 feet in length.
2. Drain height must not be higher than the original waste container inside the instrument.
Disclaimer:
Check the specifications of the central waste system for rules about discharging the
waste.
11.1 - Replacing the waste container
1. Lift the left cover and pull the waste container forwards.
2. Unscrew the cap.
3. Place the new waste container and tighten the screw cap.
4. Lift the left cover and place the waste container back into the ESR analyser unit.
Note
: If you are re-cycling waste containers, make sure that they are bleached and rinsed thoroughly.
Waste disposal
MRN-071-2-EN version 7.02 Starrsed RS
127
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...