Result is out of range, the applicable values for the
acceptable range depend on the user setting. E116 is
shown in the status line of the Sample screen and the QC
icon is blinking on the Sample screen.
ESR Result is given.
•
Try new QC sample tube (normal samples will be
finished)
•
Check acceptable range in QC settings. If results
are continuously out of range but the statistics show
identical/stable results, it should be considered to
expand the acceptable assay range with QC Settings
•
If this error persists check/clean instrument
E117: Uncorrected QC result is out of acceptable range, but corrected result is within range!
ESR Result is given.
Temperature correction not activated.
•
Consider QC Sample as correct. The mean value is
assayed with temperature correction
•
Check temperature correction setting.
E118: Uncorrected QC result is within acceptable range, but corrected result is out of range!
ESR Result is given.
Temperature correction not activated.
•
Consider QC Sample as not correct
•
Try new QC sample tube (normal samples will be
finished)
•
Check acceptable range in QC settings
•
If this error persists check/clean instrument
•
Check temperature correction setting.
QC result with ESR error
no ESR Result is given
•
Check general ESR data, see
•
Check sample tube volume
•
Try new QC sample tube
QC result with ESR warning
ESR Result is given
•
Check general ESR data, general
•
Check limit settings
Note: QC Error messages are only shown and stored in QC results and not send to LIMS.
QC result is given with the same general errors and warnings as a normal patient ESR-result
Screen messages
QC icon is blinking at Sample screen
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
169
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...