9.2.2 - Report 30 Minute mode
+ REPORT E
(Not to scale)
- StaRRsed--
Date 25/04/2019
Time:
15:28
1
2
3
4
5
6
7
8
9
10
915001
42
84
75
CLEAR
17
30
23
EDTA
9.3 - ESR Error
Error messages can be found on the printout in column 10.
If errors are found during the measurement, the ESR analyser unit will give an audible alarm.
The Error message is displayed on the main screen.
9.3.1 - ESR Error and Warning code messages
ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10.
Error: No ESR result is given
Warning: Result should be reviewed before release.
The following codes are defined:
0
No errors
1
No cells/plasma found
Error
No contents could be detected in the pipette.
2
ESR Probably > 140 mm
Error
Extremely high ESR value.
3
Too many borders found
Error
More than three borders found, possibly air
bubbles. See Section Troubleshooting
.
4
Column height <nnn>
Warning
Column height must be between 180 and
210mm. <nnn> = the actual column height.
5
Measure error
Warning
The down count is not equal to the up count
from the measure head.
6
Bubbles on top
Warning
Air bubbles on top of the ESR. See Section
Reporting
MRN-071-2-EN version 7.02 Starrsed RS
119
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...