Work instruction Number
240
Page
1 of
1
Purpose:
Cleaning diluent system
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
001, 2015
Step 1
1. Remove the suction-tube from the diluent pinch valve. Disconnect the tube and connect a new tube
2. Fill a bottle with 50 ml bleach (sodium hypochlorite) and 950 ml de-ionized water. (5% solution)
3. Place the suction tube in chlorine solution.
4. Use the
[Prime]
function to fill the dispenser system with the disinfectant.
5. After the prime sequence stops, press
[Prime]
to fill the dispenser system with the disinfectant.
6. Leave the disinfectant in the system for 15 minutes.
Step 2
1. Take the diluent suction tube out of the disinfectant.
2. Wipe the tube clean and dry with a tissue.
3. Empty the bottle and refill it with hot de-ionized water (80°C).
4. Place diluent suction tube in the bottle with hot water.
5. Use the
[Prime]
function.
6. After the prime sequence stops, press "
Prime
” again to fill the dispenser system with the hot water.
Step 3
1. Empty the bottle.
2. Clean the bottle with new hot de-ionized water (80°C)
3. Reconnect the diluent container.
4. Perform another
[Prime]
.
5. After the prime sequence stops press
[Prime]
again to fill the dispenser system with the new diluent solution.
Work instruction Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
223
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...