15.1 - Peristaltic pumps
The Peristaltic pumps are located under the Waste bottle housing assembly cover.
15.1.1 - Rinse solution not primed through the system
1. Check the rinse tube condition. It may be worn or leaking or incorrectly fitted.
Check the pick-up tube in the rinse bottle or cubitainer, it may have become detached from the tube
connector in the cap. See
Peristaltic pump tube replacement
2. Check rinse solution level.
▪ If the level is insufficient, a message is displayed and the alarm sounds!
3. One of the tubes carrying the rinse solution may be blocked or kinked.
15.1.2 - Rinse solution spilling over the instrument
If rinse solution spills over the top of the pipettes, the following items must be checked:
1. Vacuum pump:
Check the vacuum pressure: Check the airflow, go to
Maintenance
tab -
Check sensors
and select
Check flow
sensor
.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
143
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...