7
Limit error
Error
One of the following limits are out of the setting
range:
•
ESR Time
•
Column height
•
Dilution
•
Bubbles on top
•
Hazy aspect
•
Temperature
9.4 - Limit error settings
When an option (at Limit error settings) is set to
YES
and this limit error occurs, results will be printed/send
to the LIMS. Together with the sedimentation time and dilution rate (which are still printed at the usual
position), the operator/analyst can see what caused the error and may or may not use the ESR values which
are preserved in the error message.
When an option is set to
NO
and this limit error occurs, the fields for
30 min ESR, 60 min ESR and the
temperature corrected ESR are filled with spaces and thus results are not printed/send to the LIMS.
The error message in the error field (column 10) indicates that at least one of the limits (ESR time, dilution
rate, column height, bubbles on top, hazy aspects and temperature) has been exceeded.
Description of the error message
L_err(hhh/www/ttt/ccc)
:
•
L_err
means it is a "limit error"
•
hhh
is the 30 minutes ESR
•
www
is the 60 minute ESR
•
ttt
is the temperature corrected 60 minute result
•
ccc
is the column height
Example of a limit error message:
•
L_err( 42/ 84/ 75/200) means 42 mm in the 30 minute method and temperature correction 75 with a
correct column height.
•
L_err(---/ 84/ 75/200) means 84 mm in the 60 minute method and temperature correction 75 with a
correct column height.
Note:
If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it
is possible that not all expected results are send to the LIMS.
Reporting
120
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...