Note:
Deionized water
Add one or two drops of Starrsed Saline to the deionized water to avoid
Reagent alarm
.
Starrsed Diluent
The solution should be discarded if it becomes turbid.
If onboard containers are used:
If the diluent does become turbid, clean the diluent bottle thoroughly with 50 ml bleach
(containing approximately 5% Na-hypochlorite) and 950 ml de-ionized water (=5% solution).
Make sure that the container is thoroughly rinsed with de-ionized water after cleaning.
If large cubitainers are used:
Replace the Diluent cubitainer.
Starrsed RS with long level sensors:
To replace the bulk reagent containers:
1. Remove level sensors and spacers from the empty container.
2. Remove empty container.
3. Place new container.
4. Remove the cubitainer screw caps and pull the necks of the bottle packs out of the cardboard box.
5. Install the level sensors and spacers.
6. Add new reagent information in the software.
WARNING:
Wrongly placed pickup tubes causes incorrect results or instrument malfunction.
Make sure to place the appropriate level sensors in the cubitainers by checking the colored
labels/colored tubes matching the color codes on the cubitainers.
The sensors and the reagents have the following color codes:
Reagent
Color code
Starrsed Rinse solution
Green
Starrsed Saline
Yellow
General maintenance
138
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...