Note
: The vacuum is ok if the indicators are shown green (in Sample mode) or flow is in range: Flow:
0980
± 60 / Abs:
0320
±10 (In Service mode).
2. Rinse actuator:
When rinsing, the rinse actuator must be energized. The rinse actuator can be found under the top cover
at the top of pipette being rinsed. The functioning can be checked, go to
Service - Manual settings - Rinse
actuator
ON/OFF.
3. Rinse nozzle:
Must engage with the bottom of the pipette. See
.
4. Piercing pin in the rinse nozzle:
It must be straight. The piercing pin is to pierce the bottom meniscus when a filled pipette is at the rinse
position.
5. Flow from the rinse nozzle
The nozzle or tubing may be blocked.
▪ Check with
Prime Disinfectant
if sufficient disinfectant flows through the system.
15.1.3 - Rinse and Saline pump failure
Rinse pump:
1. Liquid flows back into the rinse bottle or cubitainer.
▪ Replace the rinse pump tube.
2. At the start of the rinse sequence the first pipette is not washed.
Replace the rinse pump tube, see
Peristaltic pump tube replacement
Saline pump:
1. Liquid flowing back to the saline bottle or cubitainer.
▪ Replace the saline pump tube.
2. Needle is not washed sufficiently.
Replace the saline pump tube, see
Peristaltic pump tube replacement
Note:
The wider bore tube is for the rinse pump.
The narrower bore tube is for the saline pump.
Troubleshooting
144
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...