Work instruction Number
192
Page
1 of
2
Purpose:
Fill and Clean with adapter
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, September 2016
.
Cleaning agent preparation Starrsed RS: Fill and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter VERA119002 with hot
de-ionized water (80 °C) till the first mark in the
adapter. (180 ml).
2. Add cleaning agent (QRR 010905) till the
second marker in the adapter. (18 ml).
3. Place the two caps on the adapter and mix
well.
4. Put the adapter in the Input-pool.
5. Select
Maintenance -
Prime/Clean,
button
Fill and
clean
.
Start Fill and clean procedure:
1. Select button OK.
2. The adapter will be transported to the needle position.
3. The needle goes down and the fill and clean process is started.
4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the
release position and will be transported into the End-pool.
Work instruction Starrsed RS
250
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...