E7: Outer needle motor position
error! Timeout! (piercing)
Outer needle did not go down
within a certain time limit.
•
Outer needle motor is faulty.
•
Outer needle motor driver
on needle board is faulty.
•
Outer needle is blocked.
E8: Fillnozzle not in fill position!
Fill nozzle did not reach the fill
position within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
E9: Air flow failure!
ESR analyser unit was not able to
get a stable reading during the
vacuum test before aspiration the
sample.
•
Check for leakage on the
pipette or fill nozzle.
E10: Sample probe was jammed.
Check both needles before
sampling!
Sample probe was probably
jammed when going down and
exceeded the maximum current
level.
Sample probe went back to its
home position after the error.
•
Check if outer needle is
clogged up with rubber.
•
Sample probe maybe bend.
E11: Sample probe not on position
(going down)! Timeout error!
Sample probe did not go down
within a certain time limit.
•
Sample probe motor is
faulty.
•
Sample probe motor driver
on needle board is faulty.
•
Sample probe is blocked.
E12: Dilution error: wrong or no
diluent flow. Check the diluter!
Diluter malfunction
•
Check diluent flow sensor
•
Check tubes diluter system
E13: Fillnozzle not in home
position!
Fill nozzle did not reach the Home
position within a certain time limit.
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver on
nozzle board is faulty.
•
Fill nozzle is blocked.
E14: Outer needle motor position
error! (home)
Outer needle did not reach the
home (top) sensor within a certain
time limit.
•
Check home (top) sensor.
•
Outer needle motor is faulty.
•
Outer needle motor driver
on needle board is faulty.
•
Outer needle is blocked.
Appendix for Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
187
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...