Citrate mode: When the Starrsed analyzer is used in the Citrate mode, the Starrsed Control material must be
diluted manually by transferring the necessary amount of material into a pre-citrated ESR blood collection
tube. Immediately after re-suspending, transfer the necessary amount of material into a pre-citrated tube
according instructions of the tube manufacturer. Close the tube with the mixture and invert at least 12 times,
then place the sample into the analyzer.
1.
When using LAB ID: Link the Lab ID with Starrsed Control Sample ID, see chapter Linked QC IDs.
Attach the lab ID label on the tube on top of the original Starrsed Control label
2. Rub the Starrsed Control tube between hands until packed cells have been completely re-suspended.
(See also video instruction https://portal.rrmechatronics.com/whatiseqas/)
3. Place tube for at least 15 minutes on a roller mixer or rotator mixer for thoroughly
mixing. Avoid foaming. DO NOT VORTEX.
NOTE:
To ensure consistent and reproducible results, the Control material must be thoroughly mixed and
handled in the same manner each time.
4. Place Starrsed Control tube immediately after mixing into the analyzer.
5. Start the Sample mode. The Starrsed Control sample is processed in the same manner as a patient
sample. Depending on the settings in "QC settings", a request and/or result is sent to the LIMS.
6. Restore tube after each use (at 18°-30°C).
For detailed information see the Starrsed Control Package Insert.
The contents of one tube is sufficient for three Control samples. Do not mix residual material with material
from other tubes. Do not re-use empty tubes.
The software interface is described in the chapter
Starrsed Control should be disposed of as medical waste.
10.2.6 - QC Results
The measured QC results are compared with the Assay mean value and the acceptable range. The applicable
values for the acceptable range depend on the user setting. See chapter "
" for more
information.
If applicable, the QC result is reported to LIMS using the chosen LIMS settings and limit error settings.
10.2.6.1 - QC Error messages
The general ESR errors and warnings are also applied on the QC results, see "
"
When the result is within range, no message is shown.
Quality control
MRN-071-2-EN version 7.02 Starrsed RS
125
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...