6.6.5 - QC Settings
In "QC settings" the following options can be selected:
Use default assay range
ON (=Default setting): The default assay range which is predetermined for the current batch of Starrsed
Control is used. These ranges cannot be changed in this option.
OFF:
Use custom assay range
.
Use custom assay range
ON: The lab can establish their own acceptable ranges, both ranges can be set from a minimum of +/- 2
mm to a maximum of +/- 15 mm.
OFF:
Use default assay range
.
Note: It is advised to use the default assay ranges. Use caution when setting the custom ranges. A too
narrow range may cause unjustified rejection of QC sample results, subsequent rejection of patient
results and undue burden on maintenance.
Send QC request to LIMS
▪
ON: A sample request for the QC sample is send to LIMS. The QC sample will only be processed if the
LIMS responds with YES.
Use this setting if the ESR result of the QC sample will be send to LIMS as well and LIMS requires that all
samples are requested first.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
91
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...