15.10 - Rinse nozzle (wash station) alignment errors
If there is a mechanical obstruction, the rinse nozzle may not align correctly with the pipette. Check that the
two pieces of plastic tubing connected to the rinse nozzle have enough slack to allow for movement of the
nozzle.
Technical support is needed if the rinse nozzle does not line up with the pipette.
15.11 - Hazy reports
"Hazy" reports are usually caused by build-up of proteins on the inner wall
of the pipettes. Another cause is growth of micro organisms in the diluter
system. It is extremely important that the system is kept sterile.
Erroneous haziness looks more like haemolytic plasma than 'normal' hazy
plasma. If there is an abnormal number of 'hazy' aspects, check the
dispenser system carefully and clean it if there is any doubt.
Symptoms for a contaminated instrument:
1. If more than
3… 5
out of
20
measurements are reported
HAZY.
2. Poor meniscus may be reported.
3. Incorrect results.
In case of hazy reports:
1. First run an extra Fill and Clean sequence.
2. Check after a day's run if haziness is decreased. See
3. When there are still many reports, it is recommended to clean the
diluter system. See
WI-240 Cleaning the diluent system
Example of haziness
Troubleshooting
158
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...