4. Press
Mix position
, rack is now held in the gripper.
5. Press
TUBE position TUBE#
1, mixer is now moving to the position "1".
6. Press
Read barcode
, the barcode information is shown above the key.
7. Check if the information is the same as on the barcode label.
If barcode reader errors are persisting, call for service.
It could also be possible that the barcode label is created in a not readable code (See technical
specifications)
15.23 - Tube sensor detection test
To test the tube sensor detection:
1. Select
Service - Rack unit - Position motor
.
2. Switch on
Check tube detection sensor.
3. Fill a sample rack with a some sample tubes with positions 1-3-5 filled.
4. Press
IN position
and put the rack inside the grabber.
5. Press
MIX position
, rack is now held in the grabber.
6. Press
TUBE position TUBE#
"1" (mixer is now moving to the "1" position). The sensor indicator must light
up, “
Tube detected” is shown next to the green block.
7. Press
TUBE position
TUBE#
"2", the sensor indicator must not light up, “
No tube !” is shown next to the
grey block.
8. Repeat this for all applicable tube positions.
If tube detection errors are persisting, call for service.
15.24 - Tube not in 1 position
Test the first hole settings, select
Service
-
Rack unit
-
Position motor
.
Insert a sample rack with a sample tube in the grabber.
1. Press
IN position
and put the rack inside the grabber.
2. Press
Mix position
, rack is now held in the grabber.
3. Press
TUBE position
, mixer is now moving to the 1 position.
4. Press at Outer needle
DOWN,
observe and check the alignment of the first hole.
When there is a misalignment correct the setting by turning the wheel of the positioning motor till a correct
alignment is achieved. Press
Set as tube position
and press
Save settings
.
15.25 - Temperature error
If the measured room temperature is out of range, E27: "Temperature" sensor out of range. Check
Settings !" appears.
Select
Settings
-
General settings
-
Temperature (°)
The displayed value is the temperature according the build-in thermometer.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
167
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...