Check mixer motor setting
•
Mixer motor is blocked
•
Mixer motor is faulty
•
Mixer motor driver on rack board is faulty
•
Check mixer motor belt.
E70 is the error code and the text in the status line Can't home mixer! Rack not on mix position. Mixer motor
send to home position , but position gripperr motor is not on the MIX position.
•
Check the position motor.
For error numbers and details see the
Appendix - Error list Starrsed RS
Problem:
After a crash the settings of the Mixer motor screen are changed
in the Cfg.ini file
Solution:
To restore the standard mixer settings screen with the "long mix
test" and "short mix test" buttons, open the file Cfg.ini with Notepad, lock
up the row "Rack version= " and change the value to 22. Then save and
close the file.
15.20.3 - Check the mixer
The mixer is integrated with the rack grabber and will start mixing as soon the first barcode label is read.
Check the movement during the mixing and check the mixer has enough clearance to turn.
Check the following settings in
Service - Rack unit - Mixer & conveyer motor
,
•
Number of turns long mix (default = 8). This is the first mix of a new rack.)
•
Number of turn's short mix (default = 1). This is the second, third, etc… mix of the same rack.
15.21 - LIMS Communication errors
The following error numbers E30 till E37 are related to ACK/NACK handshake communications between the
Starrsed RS and the Host Computer.
•
Check the communication cable between the Starrsed RS and Host computer
•
Check the serial port settings (Baudrate, etc)
•
Check protocol settings
•
Check Host Computer settings
15.22 - Barcode reader test
To test the barcode reader:
1. Select
Service - Rack unit - Position motor
.
2. Insert a sample rack with a sample tube in the grabber on the first position.
3. Press
IN position
and put the rack inside the gripper.
Troubleshooting
166
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...