Multi-Drop Command Coding Overview
Multi-Drop Command Coding Overview
Page
Page
F-5
F-5
CVOL nnnn.n
CVOL nnnn.n
Response: “
Response: “
RVOL
RVOL
””
Set sample volume to
Set sample volume to
nnnn.n
nnnn.n
in feet
in feet
3
3
(CF). (Causes LASAIR II to sample
(CF). (Causes LASAIR II to sample
by CF and to sample by volume.)
by CF and to sample by volume.)
Maximum value = 1439 CF
Maximum value = 1439 CF
Minimum value = 0.1 CF
Minimum value = 0.1 CF
Multi-Drop CTD Command Data Format
Multi-Drop CTD Command Data Format
This data format is identical to the MiniLaz output. The
This data format is identical to the MiniLaz output. The six LASAIR II channels
six LASAIR II channels
are converted to five channe
are converted to five channels for this report. See the “
ls for this report. See the “
CTD
CTD
” command
” command
description for additional details.
description for additional details.
R
RTD
TD
TI hh:mm:ss
TI hh:mm:ss
Time (Start of sample)
Time (Start of sample)
DA yy/mm/dd
DA yy/mm/dd
Date (Start of sample)
Date (Start of sample)
ID Loc_str
ID Loc_str
Sample Location
Sample Location
SN s/n
SN s/n
Sensor Serial Number
Sensor Serial Number
LS #
LS #
Laser Status (1 = good / 0 =
Laser Status (1 = good / 0 = bad)
bad)
FR ###.##
FR ###.##
Flow Rate (cfm)
Flow Rate (cfm)
VL ###.##
VL ###.##
Sample Volume (cfm)
Sample Volume (cfm)
C1 ########
C1 ########
Channel 1 Raw Differential Counts
Channel 1 Raw Differential Counts
C2 ########
C2 ########
Channel 2 Raw Differential Counts
Channel 2 Raw Differential Counts
C3 ########
C3 ########
Channel 3 Raw Differential Counts
Channel 3 Raw Differential Counts
C4 ########
C4 ########
Channel 4 Raw Differential Counts
Channel 4 Raw Differential Counts
C5 ########
C5 ########
Channel 5 Raw Differential Counts
Channel 5 Raw Differential Counts
TM ###.##
TM ###.##
T
Temperature (Degrees
emperature (Degrees F)
F)
RH ###.##
RH ###.##
Relative Humidity (percent)
Relative Humidity (percent)