Multi-Drop Command Coding Overview
Multi-Drop Command Coding Overview
Page
Page
F-3
F-3
CDELAY sec
CDELAY sec
Response: “
Response: “
RDELAY sec
RDELAY sec
””
This is a backward compatible command. De
This is a backward compatible command. Delay is specified as start of first
lay is specified as start of first
sample to start of next sample. Sample volume
sample to start of next sample. Sample volume or sample time must be set
or sample time must be set
first for this command to
first for this command to work properly. Ti
work properly. Times are approximate
mes are approximate if sample
if sample
volume is set rather than sample time.
volume is set rather than sample time.
Set continuous sampling on/off (1/0). If turned off, then number of
Set continuous sampling on/off (1/0). If turned off, then number of samples
samples
is set to one. Use CNOS to set
is set to one. Use CNOS to set number of samples.
number of samples.
CDT yy mm dd hh mm ss
CDT yy mm dd hh mm ss
Set date and time. Either two
Set date and time. Either two digit or four digit year is ok.
digit or four digit year is ok.
Response: “
Response: “
RDT
RDT
””
CES
CES
End Sampling.
End Sampling.
Response: “
Response: “
RES 1
RES 1
””
(success) or
(success) or
““
RES 0
RES 0
” (failure)
” (failure)
CFQ
CFQ
Response: “
Response: “
RPQ 1
RPQ 1
””
or
or
““
RPQ 0
RPQ 0
””
Flush data queue--erase all data. Note: LASAIR
Flush data queue--erase all data. Note: LASAIR II can un-erase data from
II can un-erase data from
this command if no samples are taken, and locations are not erased.
this command if no samples are taken, and locations are not erased.
CID name
CID name
Response: “
Response: “
RID name
RID name
””
Set sample ID (location) to given string. This command translates all
Set sample ID (location) to given string. This command translates all
characters to upper case. The
characters to upper case. The only legal characters in this command are
only legal characters in this command are
A-
A-
Z, a-z, and 0-9
Z, a-z, and 0-9
. Other characters which are legal for LASAIR
. Other characters which are legal for LASAIR II
II
location names are not legal here.
location names are not legal here.
CISDELAY sec
CISDELAY sec
Response: “
Response: “
RISDELAY sec
RISDELAY sec
””
Set inter-sample delay in seconds. Delay is from end-to-start.
Set inter-sample delay in seconds. Delay is from end-to-start.
CKB n
CKB n
Response: “
Response: “
RKB m
RKB m
””
Lock/unlock the keyboard (1/0). If the keyboard is not locked, the
Lock/unlock the keyboard (1/0). If the keyboard is not locked, the
instrument can be stopped or controlled from front panel.
instrument can be stopped or controlled from front panel.
CNOS nnn
CNOS nnn
Response: “
Response: “
RNOS mmm
RNOS mmm
””
Set total number of samples. Setting to -1 c
Set total number of samples. Setting to -1 causes continuous sampling.
auses continuous sampling.
If the operation is successful,
If the operation is successful,
nnn
nnn
=
=
mmm
mmm
..
CPDELAY sec
CPDELAY sec
Response: “
Response: “
RPDELAY sec
RPDELAY sec
””
Set purge delay in seconds.
Set purge delay in seconds.