7.14.2 Output On/Off Delay Function (Sequence 2: OUTPUT
Settings)
The sequence 2 group enables the OUTPut on/off delay to be controlled with
triggers.
OUTP:TRIG
Sets whether to turn the output on or off when a trigger is applied.
Command
OUTPut[:STATe]:TRIGgered {ON|OFF|1|0}
OUTPut[:STATe]:TRIGgered?
Val
u
e:
ON(1)
Turn the output no when a trigger is applied
OFF(0)
Turn the output off when a trigger is applied (default)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns whether the output is turned on or off when a trigger is applied in the NR1
form in response to OUTP:TRIG?.
TRIG:SEQ2:DEL:ON / TRIG:OUTP:DEL:ON
Sets the delay for turning on the output.
Command
TRIGger:EQuence2:DELay:ON {<numeric>|MIN|MAX}
TRIGger:SEQuence2:DELay:ON?
TRIGger:OUTPut:DELay:ON {<numeric>|MIN|MAX}
TRIGger:OUTPut:DELay:ON?
Parameter
Val
u
e:
0.0 to 10.0 0 is no delay (The default value is 0.0)
Unit
S
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the delay until the output is turned on in the NR3 form in response to
TRIG:SEQ2:DEL:ON? / TRIG:OUTP:DEL:ON?.
See
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...