PAT-T 4KW
7-19
7
Remot
e
Cont
ro
l
*SRE
Sets the service request enable register.
The service request enable register is used to select the summary messages in the
status byte register that will be able to perform service requests.
To clear the service request enable register, send *SRE 0. If the register is cleared,
service requests cannot be generated by status information.
Command
*SRE <NR1>
*SRE?
Parameter
Val
u
e:
0 to 255
An SCPI error (-222, “Data out of range”) occurs if outside the
range.
(Example)
Sending *SRE 8 sets bit 3 of the service request enable register. Each
time the summary bit (bit 3) of the QUEStionable status register in the
status byte is set, a service request message is generated.
Response
Returns the value of the service request enable register in the NR1 form in response
to *SRE?.
*STB
Queries the contents of the status byte register and the MSS (master summary
status) message.
The response is the same as serial polling only with the exception that the MSS
message appears in place of the RQS message in bit 6.
Command
*STB?
Response
Returns the value of the status byte register and the MSS message (bit 6) in NR1
form in response to *STB?.
See
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...