PAT-T 4KW
7-21
7
Remot
e
Cont
ro
l
7.8
Output Setting
Output on / off
OUTP
Turn the output on/off. The condition set by the Output on/off Delay function is
disabled.
Command
OUTPut[:STATe][:IMMediate] {ON|OFF|1|0}
OUTPut[:STATe][:IMMediate]?
Parameter
Val
u
e:
ON(1)
Output on
OFF(0)
Output off (default)
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the output status in the NR1 form in response to OUTP?.
OUTP:PON:STAT
Sets the output state at power-on.
Command
OUTPut:PON:STATe {RST|AUTO}
OUTPut:PON:STATe?
Parameter
Val
u
e:
RST
Output off at power-on (default)
AUTO
Output on at power-on
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the output state at power-on as character data in response to
OUTP:PON:STAT?.