PAT-T 4KW
7-31
7
Remot
e
Cont
ro
l
7.13 Preset Memory Function
This section explains the commands related to the memory functions of the PAT.
•
When the phase input mode is changed to the single-phase input mode from the
three-phase input mode, the setting value of the output current and the
overcurrent protection (OCP) is limited to 75 % of the rated value of output
current and the overcurrent protection (OCP) of the three-phase input mode.
MEM:SAV
Saves the present voltage and current settings of the PAT to the preset memory.
Command
MEMory:SAV {1|2|3}
Parameter
Val
u
e:
1
Preset memory A
2
Preset memory B
3
Preset memory C
MEM:RCL
Recalls a preset memory.
Command
MEMory:RCL {1|2|3}
Parameter
Val
u
e:
1
Preset memory A
2
Preset memory B
3
Preset memory C
MEM:KLOC
Sets whether preset memory values can be recalled from the panel even when the
lock function is enabled. MEM:RCL allows preset memory values to be recalled
regardless of this setting.
Command
MEMory:KLOCk {ON|OFF|1|0}
MEMory:KLOCk?
Parameter
Val
u
e:
ON (1)
Unable to recall preset memory values in the locked
condition (Default)
OFF (0)
Able to recall preset memory values in the locked
condition
For the setting that is applied when *RST is sent, see Table A-16.
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...