PAT-T 4kW
2-7
2
Inst
allat
ion and
Preparation
Possible electric shock.
A maximum current of 17 A in the three-phase input mode and 22 A
in the single-phase input mode flows in the PAT at the rated load.
When switching the mode from the three-phase input mode to the
single-phase input mode, Be sure to use the appropriate type of
power cable as specified following instruction and refer to the
connection procedure for the single-phase input mode.
(Three-phase input Mode) Connection procedure
This product does not come with a power cable. Use a four-core PVC insulated
cable with a nominal cross-sectional area of at least 3.5 mm
2
for 600 Vac.
Fig.2-4
Connecting the po
w
er cable for three-phase inp
u
t mode
(example in
w
hich the optional po
w
er cable is
u
sed)
1
Check that the AC po
w
er line meets the nominal inp
u
t rating of the
prod
u
ct.
The voltage that can be applied is any of the nominal power supply voltages in the
range of 200 Vac to 240 Vac. The frequency is 50 Hz or 60 Hz.
2
Check that the POWER s
w
itch is t
u
rned off.
3
Connect the po
w
er cable to the AC I
N
PUT terminal block as sho
w
n in
4
Attach crimping terminals to the s
w
itchboard end of the po
w
er cable.
Attach crimping terminals that comply with the terminal screws on the switchboard
5
T
u
rn off the s
w
itchboard breaker.
6
Connect the po
w
er cable by matching the U, V, W, and
(G
N
D)
terminal of the s
w
itchboard.
Connect the power cable firmly so that it does not come loose.
WARNING
GND
(GND)
W
V
U
(GND)
: Green
W
: Black
U
: Red
V
:
W
hite
M6
screws
AC INPUT terminal
b
lock
Terminal
b
lock co
v
er
Use this screw hole
(for three-phase mode)
Terminal
b
lock co
v
er
( Three-phase mode)
*
W
hen the power ca
b
le
layout is in re
v
erse
direction, this terminal
co
v
er should
b
e attached in
the up-side down position.
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...