4-26
PAT-T 4kW
■
Setting
1
Set all the req
u
ired parameters s
u
ch as the o
u
tp
u
t voltage and o
u
tp
u
t
c
u
rrent.
2
Press the LOCK (SHIFT+LOCAL) key.
The LOCK LED on the display illuminates, and the lock is enabled.
■
Release
To release the lock function, hold down LOCK (SHIFT+LOCAL) key until the
LOCK LED on the display turns off.
4.10 Switching from Remote to Local Mode
When the PAT is operating under remote control, the RMT LED on the display
illuminates.
Fig.4-23 Panel display example in remote mode
To switch from the remote mode to the local mode (panel operation) from the panel,
press the LOCAL key.
If SYST:RWL or IEEE488.0 llo is specified in remote mode, the PAT does not
switch to local mode (panel operation) even when the LOCAL key is pressed.
4.11 Remote Sensing Function
The remote sensing function is used to reduce the influence of voltage drops due to
the load cable resistance and stabilize the output voltage across the load.
The remote sensing function of the PAT can compensate up to approximately 0.6 V
for a single line. Select a load cable with sufficient current capacity so that the
voltage drop in the load cable does not exceed the compensation voltage. When the
remote sensing is performed, apply the voltage of sensing point (at the load
terminal) which does not exceed the rated output voltage. When the remote sensing
is performed at near to the maximum output voltage, the output is limited at the
maximum voltage (105 % of the rated output voltage).
To perform remote sensing, an electrolytic capacitor may be required at the sensing
point (load terminal).
EXT
EXT
LOCK
PRESET A
B
C
A
V
OUTPUT
ALARM
CC
CC
Lights on
RMT
See
See
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...