PAT-T 4kW
5-11
5
Ext
e
rnal C
ontr
o
l
5.4.2
External Resistance (Rext) Control
To control the output voltage using Rext, select the CV control source in the
CONFIG settings from the following two modes.
•
External resistance control 10 k
Ω
→
MAX OUT (CF20: 2)
The output voltage (Eo) varies in the range of 0 to the rated output voltage
(Ertg) by setting the external resistance (Rext) in the range of 0 k
Ω
to 10 k
Ω
.
Eo = Ertg
×
Rext /10 [V]
Rext = 10
×
Eo /Ertg [V]
•
External resistance control 10 k
Ω
→
0 OUT (FAIL SAFE) (CF20: 3)
The output voltage (Eo) varies in the range of the rated output voltage (Ertg) to
0 by setting the external resistance (Rext) in the range of 0 k
Ω
to 10 k
Ω
.
Eo = Ertg
×
(10-Rext)/10 [V]
Rext = 10
×
(Ertg-Eo)/Ertg [V]
•
If Rext comes loose when using the 10 kW
→
MAX OUT CV mode, excessive
voltage may be applied to the load. For safety reasons, it is recommended that
fail-safe 10 kW
→
0 OUT CV mode be used.
•
If you are using fixed resistors for Rext and controlling the output voltage by
switching through them, use a short-circuit or continuous type switch.
See
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...