5.8
External Monitoring
External monitoring of the output voltage and output current
The J1 connector consists of monitor outputs for output voltage and output current.
Table 5-4
Monitor o
u
tp
u
t of o
u
tp
u
t voltage and o
u
tp
u
t c
u
rrent
•
Shorting V MO
N
and I MO
N
to A COM can ca
u
se damage to the PAT.
•
Monitor output rating
Output impedance: 1 k
Ω
or less
Maximum output current: Approx. 10 mA
The monitor outputs are used to monitor the DC voltage (mean value).
They cannot be used to accurately monitor the AC components (ripple, transient
response, etc.) of the actual output voltage or current.
Pin
N
o.
Signal name
Description
15 and 17
A COM
Common for remote control inp
u
t
Common terminal of the o
u
tp
u
t monitor
16
I MO
N
Monitor o
u
tp
u
t of o
u
tp
u
t voltage
0 V to approx. 10 V for 0 to the rated o
u
tp
u
t voltage.
When the phase inp
u
t mode is changed to the
single-phase inp
u
t mode from the three-phase inp
u
t
mode, the setting val
u
e of the o
u
tp
u
t c
u
rrent and the
overc
u
rrent protection (OCP) is limited to 75 % of
the rated val
u
e of o
u
tp
u
t c
u
rrent and the overc
u
rrent
protection of the three-phase inp
u
t mode.
1
8
V MO
N
Monitor o
u
tp
u
t of o
u
tp
u
t c
u
rrent
0 V to approx. 10 V for 0 to the rated o
u
tp
u
t c
u
rrent
CAUTION
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...