PAT-T 4kW
4-5
4
B
a
s
ic O
p
er
a
ti
on
If the condition above occurs and you are unable to change any of the settings, turn
the POWER switch on while holding down the SHIFT key to reset the PAT to
factory default settings.
•
If the OVP/OCP settings are not appropriate
w
hen yo
u
change the load,
the load may break.
Output on/off delay functions
You can set a delay (0.1 s to 10.0 s) until the output is actually turned on or off after
you turn the OUTPUT switch on or off. This function is useful if you want to turn
the output on/off by setting a time offset according to the load characteristics.
To enable this function, set the output on delay (CF10: 0.1 to 10.0) and output off
delay (CF11: 0.1 to 10.0) of the CONFIG parameter. You can set separate delays for
on and off.
The OUTPUT LED blinks while the output on/off delay function is in operation.
Press the OUTPUT switch while the OUTPUT LED is blinking to cancel the output
on/off delay.
Press the OUTPUT switch while holding down the SHIFT key while the OUTPUT
LED is blinking to cancel the output on/off delay and forcibly switch the output.
Fig.4-6
O
u
tp
u
t on/off delay action
See
CAUTION
See
Off
Off
Blinking
On
Blinking
Delay
Delay
On
Delay
Delay
Off
Off
Blinking
Blinking
Output on
Output off
Output off
OUTPUT
LED
Output on
Cancel output on delay
Output on
b
y force
Cancel output off delay
Output off
b
y force
Output off
Delay
On
On
Blinking
Output on
Output off
Output off
OUTPUT
LED
Delay
Delay
Off
Blinking
Blinking
Output on
Output off
OUTPUT
LED
Output on
Output off
Output on
Output on
Output on
Cancel delay
Output off
Cancel delay
Output off
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...