PAT-T 4KW
Monitoring the power on/off status the the external (J1) connector
SYST:CONF:PST
Sets whether to output a low level signal at power on status or power off status when
monitoring the power on/off status externally (through the J1 connector).
Command
SYSTem:CONFigure:PSTatus {NORMal|INVerted}
SYSTem:CONFigure:PSTatus?
Parameter
Val
u
e:
NORMal
Output a low level signal while the power is on
(default)
INVerted
Output a low level signal for 10 s to 15 s when the
power is off.
Response
Returns the power on/off status as character data in response to SYST:CONF:PST?.
Queries the state of the sensing switch
SYST:CONF:RSEN
Queries the state of the sensing switch on the rear panel.
Command
SYSTem:CONFigure:RSENsing?
Response
Returns 0 if the sensing switch on the rear panel is set to off (remote sensing disabled) or
1 if set to on (remote sensing enabled) in response to SYST:CONF:RSEN?.
Displays communication error by performing a debug trace
SYST:CONF:TRAC
Sets whether to display communication errors by performing a debug trace. If the
debug trace function is turned on, error codes (example: Err-100) are shown on the
PAT display.
Command
SYSTem:CONFigure:TRACe {ON|OFF|1|0}
SYSTem:CONFigure:TRACe?
Parameter
Val
u
e:
ON(1)
Display communication errors
OFF(0)
Not display communication errors (default)
Response
Returns whether to display communication errors in the NR1 form in response to
SYST:CONF:TRAC?.
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...