7-22
PAT-T 4KW
Voltage Settings
VOLT
Sets the voltage.
Command
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<numeric>|MIN|MAX}
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
{MIN|MAX}
Parameter
Val
u
e:
0 % to 105 % of the rated output voltage
(The default value is 0 % of the rated output voltage.)
An SCPI error (-221, “Settings conflict”) occurs if
VOLT:EXT:SOUR is not set to NONE.
Unit
V
For the setting that is applied when *RST is sent, see Table A-16.
Response
Returns the voltage setting in the NR3 form in response to VOLT?. If the voltage
value is set using external input, the specified voltage is returned.
Current Settings
•
When the phase input mode is changed to the single-phase input mode from the
three-phase input mode, the setting value of the output current and the
overcurrent protection (OCP) is limited to 75 % of the rated value of output
current and the overcurrent protection (OCP) of the three-phase input mode.
CURR
Sets the current.
Command
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<numeric>|MIN|MAX}
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
[MIN|MAX]
Parameter
Val
u
e:
0 % to 105 % of the rated output current
(The default value is 105 % of the rated output current.)
An SCPI error (-221, “Settings conflict”) occurs if
CURR:EXT:SOUR is not set to NONE.
Unit
A
For the setting that is applied when *RST is sent, see Table A-16.
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...