410
Standard Adjust
Standard adjust is recommended when using stored programs 91
or 92. It
should not
be used with a user-entered calibration.
To adjust the calibration curve using the reading obtained with
the
50-mg/L standard solution, press the
SETUP
key and scroll (using
the arrow keys) to the STD setup option. Press
ENTER
to activate
the standard adjust option. Then enter
50
to edit the standard
concentration to match that of the standard used. Press
ENTER
to
complete the adjustment. See
Section 1
,
Standard Curve
Adjustmen
t for more information.
Method Performance
Precision
In a single laboratory, using a standard solution of 50 mg/L
sulfate and two representative lots of powder pillows with the
instrument, a single operator obtained a standard deviation of
±0.5 mg/L sulfate.
In a single laboratory, using a standard solution of 50 mg/L
sulfate and two representative lots of AccuVac Ampuls with the
instrument, a single operator obtained a standard deviation of ±3
mg/L sulfate.
Estimated Detection Limit (EDL)
The EDL for program 91 is 4.9 mg/L SO
4
and the EDL for
program 92 is 3 mg/L SO
4
. For more information on derivation
and use of Hach’s estimated detection limit, see
Section 1
.
Interferences
The following interfere at levels above those concentrations
listed:
Summary of Method
Sulfate ions in the sample react with barium in the SulfaVer 4
Sulfate Reagent to form insoluble barium sulfate. The amount of
turbidity formed is proportional to the sulfate concentration. The
SulfaVer 4 also contains a stabilizing agent to hold the precipitate
in suspension.
Calcium: 20,000 mg/L as CaCO
3
Magnesium: 10,000 mg/L as CaCO
3
Chloride: 40,000 mg/L as Cl
-
Silica: 500 mg/L as CaCO
3
SULFATE,
continued
Содержание DR/850
Страница 2: ...2...
Страница 8: ...8...
Страница 9: ...9 Sample Procedure Explained...
Страница 10: ...10 Sample Procedure Explained continued...
Страница 11: ...11 Sample Procedure Explained continued...
Страница 12: ...12 Sample Procedure Explained continued...
Страница 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Страница 69: ...69 SECTION 4 PROCEDURES...
Страница 70: ...70...
Страница 92: ...92...
Страница 100: ...100...
Страница 108: ...108...
Страница 116: ...116...
Страница 139: ...139...
Страница 140: ...140...
Страница 168: ...168...
Страница 176: ...176...
Страница 184: ...184...
Страница 194: ...194...
Страница 202: ...202...
Страница 214: ...214...
Страница 234: ...234...
Страница 242: ...242...
Страница 264: ...264...
Страница 270: ...270...
Страница 284: ...284...
Страница 298: ...298...
Страница 314: ...314...
Страница 324: ...324...
Страница 332: ...332...
Страница 336: ...336...
Страница 340: ...340...
Страница 354: ...354...
Страница 370: ...370...
Страница 378: ...378...
Страница 386: ...386...
Страница 398: ...398...
Страница 404: ...404...
Страница 412: ...412...
Страница 424: ...424...
Страница 436: ...436...
Страница 446: ...446...
Страница 456: ......
Страница 459: ......
Страница 460: ...460...