Summary of Method
Ammonia compounds combine with chlorine to form
monochloramine. Monochloramine reacts with salicylate to form
5-aminosalicylate. The 5-aminosalicylate is oxidized in the
presence of a sodium nitroprusside catalyst to form a blue-
colored compound. The blue color is masked by the yellow color
from the excess reagent present to give a final
green-colored solution.
Pollution Prevention And Waste Management
The ammonia salicylate reagent contains sodium
nitroferricyanide. Cyanide solutions are regulated as hazardous
wastes by the Federal RCRA. Collect cyanide solutions for
disposal as reactive (D001) waste. Be sure cyanide solutions are
stored in a caustic solution with pH >11 to prevent release of
hydrogen cyanide gas. See
Section 3
for further information in
proper disposal of these materials.
REQUIRED REAGENTS
Cat. No.
AmVer Reagent Set for Nitrogen, Ammonia, Low Range TNT (25 tests)................... 26045-45
Includes: (1) 23952-66, (1) 23954-66, (1) 272-42,
*
(50) AmVer Low Range Vials
Quantity Required
Description
Per Test
Unit
Cat. No.
AmVer Diluent Reagent, Low Range Test ‘N Tube ... 2 vials ............... 50/pkg....................... *
Salicylate Reagent Powder Pillows, 5 mL sample .... 2 pillows ............. 50/pkg.......... 23952-66
Cyanurate Reagent Powder Pillows, 5 mL sample.... 2 pillows ............. 50/pkg.......... 23954-66
Sulfide
1.
Measure about 350 mL of sample in a 500 mL erlenmeyer flask.
2.
Add the contents of one Sulfide Inhibitor Reagent Powder Pillow. Swirl to mix.
3.
Filter the sample through a folded filter paper.
4.
Use the filtered solution in step 4.
Other
Less common interferences such as
hydrazine
and
glycine
will cause intensified
colors in the prepared sample.
Turbidity
and
color
will give erroneous high
values. Samples with severe interferences require distillation. Hach recommends
the distillation procedure using the Hach General Purpose Distillation Set. See
Optional Apparatus at the end of this procedure.
Interfering Substance
Interference Level and Treatment
* Not available separately.
NITROGEN, AMMONIA, Low Range, Test 'N Tube,
continued
Содержание DR/850
Страница 2: ...2...
Страница 8: ...8...
Страница 9: ...9 Sample Procedure Explained...
Страница 10: ...10 Sample Procedure Explained continued...
Страница 11: ...11 Sample Procedure Explained continued...
Страница 12: ...12 Sample Procedure Explained continued...
Страница 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Страница 69: ...69 SECTION 4 PROCEDURES...
Страница 70: ...70...
Страница 92: ...92...
Страница 100: ...100...
Страница 108: ...108...
Страница 116: ...116...
Страница 139: ...139...
Страница 140: ...140...
Страница 168: ...168...
Страница 176: ...176...
Страница 184: ...184...
Страница 194: ...194...
Страница 202: ...202...
Страница 214: ...214...
Страница 234: ...234...
Страница 242: ...242...
Страница 264: ...264...
Страница 270: ...270...
Страница 284: ...284...
Страница 298: ...298...
Страница 314: ...314...
Страница 324: ...324...
Страница 332: ...332...
Страница 336: ...336...
Страница 340: ...340...
Страница 354: ...354...
Страница 370: ...370...
Страница 378: ...378...
Страница 386: ...386...
Страница 398: ...398...
Страница 404: ...404...
Страница 412: ...412...
Страница 424: ...424...
Страница 436: ...436...
Страница 446: ...446...
Страница 456: ......
Страница 459: ......
Страница 460: ...460...