background image

GE Healthcare
CARESCAPE Monitor B450 
Technical Manual

Software Version 2
Hardware Version B450- 01

 

All specifications subject to change without notice.

 

English
 

2062973-003 paper

September 19, 2016

© 2013, 2016 General Electric Company. All rights reserved.

Содержание CARESCAPE Monitor B450

Страница 1: ... Monitor B450 Technical Manual Software Version 2 Hardware Version B450 01 All specifications subject to change without notice English 2062973 003 paper September 19 2016 2013 2016 General Electric Company All rights reserved ...

Страница 2: ...mation in this manual applies to the software and hardware versions listed on the first page of this manual Due to continuing product innovation specifications in this manual are subject to change without notice ...

Страница 3: ...ols 6 3 System overview 9 3 1 System introduction 9 3 2 System components 9 3 2 1 Monitor 9 3 2 2 Software 10 3 2 3 Input devices 10 3 2 4 Acquisition modules 12 3 2 5 CARESCAPE Network MC or S 5 Network 12 3 2 6 CARESCAPE Network IX 13 3 2 7 Unity Network ID connectivity device 14 3 2 8 Secondary display 14 3 2 9 Printers and recorders 15 3 2 10 Service Interface 16 3 3 Controls and connectors 17...

Страница 4: ...talling a battery into the patient monitor and the PDM module 55 6 1 1 Testing the battery charge 55 6 1 2 Installing the battery into the patient monitor 56 6 1 3 Installing the battery into the PDM module 56 6 2 Mounting the patient monitor 57 6 3 Connecting a secondary display 57 6 3 1 Connections to D15K and D19KT displays 58 6 4 Installing parameter modules 60 6 4 1 Installing a PSM or a PDM ...

Страница 5: ...13 Setting the host asset number 86 7 14 Changing passwords 86 7 15 Restarting the patient monitor 87 7 16 Setting up the remote service 87 7 16 1 Configuring the remote service 87 7 16 2 Enabling remote service agent connection 88 7 16 3 Testing the connectivity 89 7 17 Transferring settings from one patient monitor to other patient monitors 89 7 17 1 Saving settings 89 7 17 2 Loading settings 90...

Страница 6: ... Network 113 8 3 10 Wireless LAN 113 8 3 11 IX printers 114 8 3 12 iPanel connection 115 8 3 13 Insite with EXC 115 8 3 14 Test completion 115 9 Theory of operation 117 9 1 Power Supply Power management subsystem 118 9 1 1 AC DC Power Supply 118 9 2 DC DC Board 119 9 2 1 PUIC controller and software 120 9 2 2 DC DC Subsystem 120 9 2 3 User Interface Controller UIC subsystem 121 9 3 CPU assembly 12...

Страница 7: ...ns 136 10 4 3 Testing the battery charge 136 10 4 4 Charging a battery 136 10 4 5 Conditioning a battery 137 10 4 6 Replacing a battery 137 10 4 7 Battery recycling 137 11 Troubleshooting 139 11 1 Visual inspection 139 11 2 Webmin Information tab 139 11 2 1 Configuration information 140 11 2 2 Device information 141 11 3 Webmin Diagnostics tab 142 11 3 1 Hardware statistics 143 11 3 2 Ping a TCP I...

Страница 8: ...bly from the rear unit assembly 183 12 2 7 Detaching the uDOM 187 12 2 8 Detaching the CPU assembly FRU 188 12 2 9 Replacing the AC DC power supply unit FRU 192 12 2 10Replacing the WLAN assembly FRU 193 12 2 11Detaching the front unit assembly from the mid frame assembly 195 12 2 12Replacing LCD display unit FRU 198 12 2 13Replacing the Speaker unit FRU 199 12 2 14Detaching the user interface boa...

Страница 9: ...mpatibility information including electromagnetic compatibility See the user s manual for the instructions necessary to operate the device safely in accordance with its function and intended use 1 2 Intended audience of the manual This manual is intended for service representatives and technical personnel who install maintain troubleshoot or repair this device 1 3 Conventions used in this manual W...

Страница 10: ... Start End case menu in OR and PACU software 1 4 Illustrations and names This manual uses illustrations as examples only Illustrations in this manual may not necessarily reflect all system settings features configurations or displayed data Names of persons institutions and places and related information are fictitious any similarity to actual persons entities or places is purely coincidental 1 5 O...

Страница 11: ...party trademarks Masimo SET is a trademark of Masimo Corporation PiCCO is a trademark of Pulsion Medical Systems SE WMM WPA and WPA2 are trademarks of Wi Fi Alliance Covidien BISx Bispectral Index BIS and Nellcor are trademarks of a Medtronic company 1 8 Responsibility of the manufacturer GE is responsible for the effects on safety reliability and performance of the equipment only if Assembly oper...

Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...

Страница 13: ...ctions The use of accessory equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system Consideration relating to the choice shall include use of the accessory in the patient vicinity and evidence that the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601 1 and or I...

Страница 14: ... instructions DANGER Shock hazard Dangerous voltage To reduce the risk of electric shock do not remove cover Refer servicing to qualified service personnel This symbol is identified by a yellow background black triangular band and a black symbol Electrostatic sensitive device Connections should not be made to this device unless ESD precautionary procedures are followed Non ionizing electromagnetic...

Страница 15: ...ation to the patient including direct cardiac application Type CF IEC 60601 1 defibrillator proof protection against electric shock Isolated floating applied part suitable for intentional external and internal application to the patient including direct cardiac application Safety ground Remove power cord from the mains source by grasping the plug Do not pull on the cable ...

Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...

Страница 17: ...ation with the other subsystems and external interfaces The main software and all platform and clinical settings are stored in a detachable flash memory uDOM The patient monitor supports two optional multi parameter hemodynamic modules Patient Data Module PDM and Patient Side Module PSM In addition it supports one single width E series parameter module that extends the monitoring capability to oth...

Страница 18: ...ighly configurable and provides many monitoring possibilities with a flexible software licensing model The monitor supports care area specific software packages for OR PACU ICU ED and NICU Each dedicated software package provides a comprehensive feature set for the different monitoring needs and can be further extended with the optional feature licenses Software license model supports trial licens...

Страница 19: ...ted to the monitor or display via one of the USB connectors The mouse allows you to select any on screen items without a Trim Knob control or a touchscreen display Remote control The remote control provides all patient monitor controls on a portable component with a Trim Knob control The remote control is connected to the patient monitor via one of the USB connectors Barcode reader The barcode rea...

Страница 20: ...n manual for a list of compatible acquisition devices and to the patient monitor s user s manual for a list of parameters each module measures 3 2 5 CARESCAPE Network MC or S 5 Network The patient monitor is compatible both with the CARESCAPE Network MC and the S 5 Network infrastructures The optional WLAN support enables wireless network communication using IEEE 802 11a b g Refer to the patient m...

Страница 21: ...rix server for iPanel connections and to the IX printers It also enables centralized Webmin access for service personnel from within the hospital and the InSite with ExC remote service connectivity to GE s support center Refer to the CARESCAPE Network Configuration Guide for details on configuring the CARESCAPE Network The iPanel application viewable from one of the monitor s display screens gives...

Страница 22: ...pported interfaces include anesthesia machines ventilators gas analyzers continuous cardiac output devices pulse oximeters transcutaneous monitors and point of care blood gas monitors Refer to the Unity Network Interface Device ID Operator s Manual and the patient monitor s supplemental information manual for a list of compatible peripheral devices and to the patient monitor s user s manual for th...

Страница 23: ...ulation results and care reports Refer to the patient monitor s user s manual for more information about printing The patient monitor supports printing to a laser printer that is connected to the patient monitor via the IX Network or directly to the IX connector in the patient monitor to a laser printer that is connected to a CIC Pro Clinical Information Center on the MC Network to a laser printer...

Страница 24: ...he other way to access Webmin locally is from a configured service PC that is connected to the patient monitor with an Ethernet crossover cable Remote access to Webmin The user also can access Webmin remotely using a configured service PC over the IX Network InSite with ExC InSite with ExC provides a set of software applications to manage diagnose and track systems at customer sites by using the I...

Страница 25: ...larm paused off area blue 7 Alarm light area blue yellow or red 8 Mains voltage indicator the green LED is lit when the patient monitor is connected to AC mains 9 Battery use indicator the green LED is lit when the patient monitor is operating on battery power 10 Battery charging failure indicator the orange LED is lit when the patient monitor battery is charging and flashing in case of battery fa...

Страница 26: ... cover 3 Receptacle for power cord and fuse holder 4 Defibrillation synchronization connector 5 Service cover 6 Recorder 7 Module slot optional NOTE The Defibrillation synchronization connector can be used only with E PSM P modules PDM module has its own defibrillation synchronization connector ...

Страница 27: ...1 Power LEDs for troubleshooting 2 Ventilation holes 3 Slide mount with connector for PDM 4 Cable clamp for power cord 5 Equipotential connector 6 Connectors for input output devices and networking 7 Device label 8 Slide mount with connector for PSM ...

Страница 28: ...equipment under warranty voids that warranty It is the user s responsibility to report the need for service to GE or to one of their authorized agents Failure on the part of the responsible individual hospital or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards Regular maintenance irrespective of usage i...

Страница 29: ...here the first three characters represent Product Code 11 GS1 application identifier for the manufacturing date of the device 151230 Manufacturing date year month day YYMMDD Note that for some product types the production identifier can have other elements instead of the ones listed above 10 GS1 application identifier for the batch or lot number followed by the batch or lot number 17 GS1 applicati...

Страница 30: ...an access control system In the clinical and Webmin applications there are certain features that require user authentication To access these features the user must log into the clinical and Webmin applications with a valid username and password Authorization Authorization is the process of granting and revoking access to information and is another key element in an access control system Although p...

Страница 31: ...ugh GE s vigorous software verification and validation process Any software update needs will be communicated by GE 3 5 2 Security operations Network security GE requires that the MC port of the patient monitor be connected to a physically or virtually dedicated CARESCAPE Network MC or S 5 Network isolated from all other networks GE requires that the IX port of the patient monitor be connected to ...

Страница 32: ...D to the wireless network The wireless MC Network shall be logically separated to its own virtual LAN with a dedicated SSID The S 5 Wireless Network shall have dedicated access points that are physically isolated from all other networks Security Authentication and Confidentiality The data transmitted on the wireless network can be secured by one of the following methods WEP 64 bit NOTE S 5 only WE...

Страница 33: ... more of the devices Bell cancel Audio off Audio pause Temporary audio off General alarm Fuse Replace with identical type and rating fuse Do not reuse Battery monitor The flashing orange symbol indicates that there is a battery failure missing battery Battery monitor The solid orange symbol indicates that the battery is being charged Battery monitor The solid green symbol indicates that the monito...

Страница 34: ...monitor Located on the battery slot cover Battery PDM Communication PDM Power indicator PDM On standby button Standby or power indicator Signal power input Signal power output Signal power input output combined ON Power connection to the mains Power supply connector The following symbols appear on one or more of the devices ...

Страница 35: ...nector for digital or analog source Color display connector port Color video input Video input connector for digital or analog source Color video output digital Video output for analog source Color video output Video output for digital source USB connector Ethernet connectors The following symbols appear on one or more of the devices ...

Страница 36: ...armful ingress of water Components not marked with an IPXn code are rated as Ordinary no protection against fluid ingress All other IPXn rated components have the degree of protection per the n rating Latex free D lite Pedi lte Add date Home Return to the main display Alternating current Green symbol on the B650 and B450 monitor front panel the monitor is being used on mains power Direct current T...

Страница 37: ...entifier Stacking limit by number Date of manufacture This symbol indicates the date of manufacture of this device The first four digits identify the year and the last two digits identify the month Manufacturer address and date of manufacture The first four digits identify the year the following two digits identify the month and the last two digits identify the day The following symbols appear on ...

Страница 38: ...breviation for label part number Identifies the device type Catalogue or orderable part number Device serial number Device hardware version Every device has a unique marking for identification The UDI marking appears on the device label Mass of typical portable RGM respiratory gas monitor configuration The indicated mass 12 kg in this example varies per RGM configuration Atmospheric pressure limit...

Страница 39: ...Lithium ion Mercury This product consists of devices that may contain mercury which must be recycled or disposed of in accordance with local state or country laws Within this system the backlight lamps in the patient monitor display contain mercury European authorized representative European Union Declaration of Conformity Indicates that the product is certified for both the U S and Canadian marke...

Страница 40: ...cates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information products will not result in any severe environmental pollution any bodily injury or damage to any assets The unit of the period is Year In order to maintain the declared EFUP...

Страница 41: ...ina only China Ministry of Industry and Information Technology identification number for Radio Transmission Equipment Type Approval Australia only The product complies with the applicable Australian standard and establishes a traceable link between the equipment and the manufacturer importer or their agent responsible for compliance Japan only Approved under Japan TELEC requirements Brazil only Ap...

Страница 42: ...tone volume Audio alarms off indicator Displays in the upper left corner of the alarm area when physiological audible alarms are turned off Audio alarms paused indicator with countdown timer Indicates all audio alarms are paused and the amount of time remaining for the alarm pause period displays as a countdown timer Displays in the upper left corner of the screen Pause audio alarms Selectable fro...

Страница 43: ...rity column of the selected alarm in the Alarm Setup Alarm Priorities Reminder volume icon Adjust the volume of the tone that sounds every two minutes when audio alarms are turned off Touch volume icon Adjust the volume of the tone that sounds when a user touches a touchscreen display Home icon Close all menus applications displayed on the patient monitor Locking setting indicator Indicates this s...

Страница 44: ...n time left Monitor battery is charging There is a white running bar inside the symbol Monitor battery failure indicator Indicates a missing battery or a battery failure PDM battery charging indicator Indicates that the PDM battery is charging PDM battery gauge indicator Indicates the charge level of the battery PDM battery failure indicator Indicates the battery is not available for use Snapshot ...

Страница 45: ...ck Completed NIBP volume icon Adjust the volume of the tone that sounds when an NIBP measurement result is available Manual NIBP icon Start a manual NIBP measurement Nellcor OxiMax SatSeconds indicator Indicates the amount of time the SpO2 saturation is outside the limits before alarms are generated SpO2 signal strength indicator Indicates the signal strength with three asterisks indicating the st...

Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...

Страница 47: ...vice PC that is connected to the IX port of the patient monitor with an Ethernet crossover cable Remote Webmin You can access Webmin remotely from a configured service PC that is connected to the patient monitor over the IX Network Requirements for service PC Network board with Ethernet port and TCP IP network installed Internet Explorer V6 or later 4 1 Local access to Webmin using the integrated ...

Страница 48: ...not touch the patient and any part of non medical equipment at the same time Some examples of non medical equipment are laser printers and non medical computers 1 Connect a service PC to the IX connector on the patient monitor using a crossover cable 2 In patient monitor select Monitor Setup Service The local browser opens and displays the Login to Webmin dialog box 3 Record the IX IP address of t...

Страница 49: ...nue to 4 4 Login to Webmin Closing Webmin 1 Restore the patient monitor s original IX Network configuration and service PC s network settings if they were changed 2 Disconnect the crossover cable from the patient monitor and from the service PC 3 Reconnect the patient monitor back to the IX Network if applicable 4 3 Remote access to Webmin using a service PC over the IX Network 1 Connect a service...

Страница 50: ...ork NOTE For more information on how to configure the IP address refer to the PC s documentation 5 Launch a web browser on the service PC 6 In the Address field type https IX IP address 10000 and press Enter NOTE IX IP address is the IX Network IP address of the patient monitor The Login to Webmin dialog box displays 7 Continue to 4 4 Login to Webmin Closing Webmin 1 Restore the original network s...

Страница 51: ...iomed Password Change space Me NOTE Username and password are case sensitive NOTE Change Me is the factory default password for the username biomed Refer to section 7 14 Changing passwords for details on how to change the default password The Webmin application opens and defaults to the Information tab Figure 1 Webmin user interface when accessed using the integrated browser ...

Страница 52: ...igure 2 Webmin user interface when accessed using the service PC 4 5 Webmin configuration modules The Webmin service interface includes the following configuration modules Select Help for additional information related to each Webmin module ...

Страница 53: ... saved to the permanent memory of the related acquisition module and the settings travel with the module from one patient monitor to another Refer to the Module Frames and Modules Technical Manual for detailed information how to change these settings National Requirements to activate France specific defaults for the ECG HR adjustment range and the reminder beep behavior Network to select and confi...

Страница 54: ... patient monitor and the connected peripheral devices Settings to transfer platform and or clinical settings from one patient monitor to another to take backup copies of the settings to an external device and to restore the settings from an external device Software Management to update patient monitor and parameter module software Time to set the date and time settings Unit and bed name to configu...

Страница 55: ...TCP IP network device Use this Webmin feature to verify connectivity with a network device on the MC Network and IX Network Log files The patient monitor collects information about different system events and errors to log files These log files help troubleshooting problems in the patient monitor and the connected peripheral devices WLAN diagnostics Use this Webmin feature for troubleshooting WLAN...

Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...

Страница 57: ... a list of compatible network and bedside devices Refer to the patient monitor s supplemental information manual for a list of compatible supplies and accessories Refer to the patient monitor s supplemental information manual and Unity Network Interface Device ID Operator s Manual to see compatible peripheral devices WARNING BEFORE INSTALLATION Compatibility is critical to safe and effective use o...

Страница 58: ...work 5 3 2 Wireless MC Network The wireless MC Network infrastructure shall be installed according to the CARESCAPE Wireless Network Configuration Guide Ensure that the wireless coverage area is adequate for the installation Refer to the sections 7 2 2 Selecting and configuring CARESCAPE Network 7 2 4 Configuring WLAN and 7 4 Setting unit and bed name to see the configuration information you need ...

Страница 59: ...are is properly installed prior to the installation of the patient monitor Mounting hardware for the patient monitor either for a stand alone installation or for an installation to an anesthesia machine or to a ventilator Mounting hardware for the PSM module Mounting hardware for the PDM module Mounting hardware for the displays Mounting hardware for the Unity Network ID connectivity device NOTE R...

Страница 60: ...ion which affords sufficient ventilation The ventilation openings of the device must not be obstructed EMI RFI interference WARNING Do not use the monitor in high electromagnetic fields for example during magnetic resonance imaging WARNING Other equipment may interfere with the system even if that other equipment complies with CISPR emission requirements CAUTION EMC Magnetic and electrical fields ...

Страница 61: ...ment near the system may cause unexpected or adverse operation of this device system Consult qualified personnel regarding device system configuration The patient monitor should be isolated from sources of strong electromagnetic and radio frequency interference NOTE Refer to the patient monitor s supplemental information manual for more information ...

Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...

Страница 63: ...module housing or frames 6 1 Installing a battery into the patient monitor and the PDM module The batteries for the patient monitor and the PDM module if included are shipped separately and need to be installed and fully charged prior to taking into use WARNING PHYSICAL INJURY Make sure the battery is completely inserted and the battery door is completely closed Falling batteries could seriously o...

Страница 64: ...or WARNING The patient monitor must always be used with a battery inserted This will ensure the functioning of the monitor and prevent data loss during possible supply mains interruptions 1 Open the battery cover by pressing the battery cover release latch down and pulling the battery door open 2 Insert the batteries one at time with the test indicator facing to front and the connector end first a...

Страница 65: ...ording to the installation instructions included with the mounting hardware WARNING Never install equipment above the patient WARNING Use only manufacturer specified mounts WARNING To prevent liquids from entering the monitor do not tilt the monitor more than 15 degrees CAUTION The device system should not be used adjacent to or stacked with other equipment Consult qualified personnel regarding de...

Страница 66: ...r with a non medical grade display which is IEC 60950 rated or equivalent meets UL and IEC specifications if a medical grade isolation is used If a non medical grade display is to be used the configuration must meet the IEC 60601 1 1 standard Refer to IEC 60601 1 1 for requirements if using non medical grade displays in the patient environment 6 3 1 Connections to D15K and D19KT displays The follo...

Страница 67: ...042768 001 Standard USB A to USB B cable up to 5 meters In the following example two 5 meter USB extenders plus a standard 5 meter USB cable extend the remote display up to 15 meters from the patient monitor USB connection 1 Connect the Type A plug of the 1st USB extender to one of the downstream ports Type A USB port on the back of the patient monitor 2 Connect the Type A plug of the 2nd USB exte...

Страница 68: ...identical measurements WARNING Ensure that the CARESCAPE respiratory modules are in vertical position when used Tilting them may result in erroneous readings 6 4 1 Installing a PSM or a PDM module You can install a PSM or a PDM module either to the patient monitor or to a remote mounting solution Connecting a PSM or a PDM module to the patient monitor 1 Connect a module by aligning it with the ins...

Страница 69: ... inserted and the battery door is completely closed Falling batteries could seriously or fatally injure neonatal or other vulnerable patients WARNING PHYSICAL INJURY Do not install the PDM above a patient Leaks from the battery cells can occur under extreme conditions The liquid is caustic to the eyes and skin If the liquid comes in contact with eyes or skin flush with clean water and seek medical...

Страница 70: ...wer supply circuit Devices which are not connected to the same circuit must be electrically isolated when operated CAUTION POWER REQUIREMENTS Before connecting the device to the power line check that the voltage and frequency ratings of the power line are the same as those indicated on the device s label If this is not the case do not connect the system to the power line until you adjust the devic...

Страница 71: ... your GE Healthcare service representative to verify or update your Aware Gateway configuration WARNING EXCESSIVE LEAKAGE CURRENT Only devices that are specified compliant with IEC 60950 1 or IEC 60601 1 may be connected to the Ethernet MC or IX ports The patient monitor can be connected either to the wired MC Network or S 5 Network Connect the MC Network or the S 5 Network patch cable to the netw...

Страница 72: ...6 9 Connecting USB devices NOTE The USB connection to a D15K or D19KT secondary display reserves one of the USB connectors in the rear panel of the patient monitor Connect the following devices to the USB ports in the rear panel of the patient monitor or at the bottom of a D15K or D19KT display Alphanumeric keyboard Mouse Remote control Barcode reader NOTE Always use a GE supplied barcode reader f...

Страница 73: ...nector with a crossover cable 1 Connect the Ethernet cable to the IX connector labelled as IX in the rear panel of the patient monitor 2 Connect the other end of the Ethernet crossover cable to the connector in the laser printer NOTE If the local printer is used within the patient environment it must always be powered from an additional transformer providing at least basic isolation NOTE Refer to ...

Страница 74: ...the multiple socket outlet protective earth conductor WARNING Non medical equipment does not provide the same level of protection against electrical shock Do not touch the patient and any part of non medical equipment at the same time Some examples of non medical equipment are laser printers and non medical computers ...

Страница 75: ...r replacing the uDOM or the touchscreen sensor Calibrate the connected touchscreens one by one as follows 1 Select Monitor Setup Service Calibrations 2 In the Enter Password dialog box type the username and password and press Enter Username biomed Password Change space Me NOTE Username and password are case sensitive NOTE The factory default password for the username biomed is Change Me 3 In the S...

Страница 76: ...e patient monitor may be configured to the MC Network or IX Network or both NOTE The following IP addresses are reserved and not valid 172 16 254 254 172 16 255 255 172 18 254 254 172 18 255 255 192 168 250 x 192 168 252 x 192 168 253 x 192 168 254 1 and 192 168 254 2 5 In the Network Configuration window below MC Network a Enter a Static IP address b Enter a valid Netmask level c Select the appli...

Страница 77: ...Network select DHCP or Manual Configuration a Enter a Static IP address b Enter a valid Netmask level c Enter a valid Default Gateway d Enter valid DNS Server addresses if applicable 3 Select the applicable Speed Duplex option 4 Select Save The S 5 virtual ID change will be applied immediately All other network configurations will be saved and become active after the patient monitor is restarted 7...

Страница 78: ...r multipath interference When set to Enabled Antenna Diversity monitors the signal from each antenna and automatically switches to the one with the better signal 5 Configure the appropriate WLAN settings Note that the configurable attributes and available values are different for MC Network and S 5 Network ...

Страница 79: ...ireless Network the frequency band is fixed to 2 4 GHz and it supports 802 11b protocol only RTS Threshold Configure the RTS Threshold value Use the default RTS Threshold value unless otherwise specified in the wireless network design A valid RTS Threshold is a numeric value within the range of 64 to 2300 NOTE This option is not configurable in the S 5 Wireless Network Fragmentation Threshold Conf...

Страница 80: ...re the DSCP settings for the following network traffic Real Time Clinical Traffic Non Real time Clinical Traffic Non Real time Non Clinical Traffic Valid DSCP ranges Real Time Clinical Data 32 47 Waveforms Parameters etc Non Real time Clinical Data 0 7 Trends Full Disclosure Printing etc Non Real time Non Clinical Data 8 23 InSite Service traffic etc DSCP Differentiated Services Code Point setting...

Страница 81: ...infrastructure NOTE The authentication method used with WEP is Open System Authentication NOTE The wireless MC Network supports only confidentiality methods None WEP 128 bit WPA PSK TKIP and WPA2 PSK AES CCMP NOTE The S 5 Wireless Network supports only confidentiality methods None WEP 64 bit and WEP 128 bit Key Index Select the Key Index to be used The Key Index of the wireless client must match t...

Страница 82: ...work 1 Log in to Webmin 2 Select Configuration Time Pass Phrase Key Enter the Pass Phrase Key for the used confidentiality method as an ASCII or HEX string as selected above WEP 64 bit WEP 128 bit Enter a 5 13 character long ASCII string The system will generate a 10 26 character hexadecimal security key and assign it to the currently selected Key Index OR Enter a 10 26 character long hexadecimal ...

Страница 83: ... Webmin 2 Select Configuration Unit and Bed Name 3 In the Unit and Bed Name Configuration window view or set the unit name and bed name for the device NOTE Use only capital letters A Z numbers dash asterisk and space The unit name may be up to seven characters long and bed name up to five characters long 4 Select Save The unit and bed name configuration takes effect immediately 7 5 Configuring pri...

Страница 84: ...ly 7 5 2 Deleting a printer NOTE Before deleting a laser printer check Monitor Setup Printing Devices Status If a printout is assigned to the printer to be deleted redirect the printout to another valid printer 1 Log in to Webmin 2 Select Configuration Printers 3 In the Sub Modules for Printers menu select Delete Printer 4 In the Delete Printer window select the printer to delete 5 Select Save The...

Страница 85: ...ial Program Initial Citrix published application that will be launched for example MUSE This field is mandatory if the server address has been entered The maximum length is 128 characters Citrix Session Timeout in Minutes Valid values are between 0 99 Keeping the session alive allows users to quickly return to their last session but locks a Citrix license Selecting 0 disables the timeout feature U...

Страница 86: ...tal defined prefix that is used as the first two characters in a temporary patient ID to ensure its uniqueness inside the hospital 1 Log in to Webmin 2 Select Configuration Admit Settings 3 In the Sub Modules for Admit Settings menu select Patient ID Prefix 4 In the Patient ID Prefix window enter a 2 character prefix NOTE Valid values are uppercase letters and numbers 5 Select Save All changes tak...

Страница 87: ...n 7 8 3 Configuring Length Delimited Parser information Points to note If you configure Age you must either select the Age Unit item or one of the age units e g Years Months Weeks Days below Fixed Option If you configure Height you must either select the Height Unit item or one of the height units e g Feet Inches Meters Centimeters Millimeters below Fixed Option If you configure Weight you must ei...

Страница 88: ...arser information In this example the barcode contains 10 items The following table lists the starting position and length of each item The following sample shows the corresponding entries in the Admit Discharge Configuration window Item Item selection in the top part of the screen Fixed Option selection Height Unit Both Height and Height Unit Non Fixed Height only Select value from drop down list...

Страница 89: ...Height you must either select the Height Unit item or one of the height units e g Feet Inches Meters Centimeters Millimeters below Fixed Option If you configure Weight you must either select the Weight Unit item or one of the weight units e g Kilograms Grams Micrograms Pounds Ounces below Fixed Option For an example of admit discharge configuration for a length delimited parser see Example of char...

Страница 90: ...barcode it will be misinterpreted as a field delimiter 3 Below Gender Code enter the codes that identify Male and Female 4 Below Fixed Option select the applicable value 5 Scroll to the bottom of the window and select Save All changes take effect immediately Example of character delimited parser information entry In the following example the barcode contains 12 items and uses the pound sign as a d...

Страница 91: ...hows the corresponding entries in the Admit Discharge Configuration window 7 8 5 Barcode data specifications Points to note The maximum length of the entire barcode is 300 Height 8 Height Unit 9 Weight 10 Item Sequence number of the item in the barcode ...

Страница 92: ... Day of Birth and Year of Birth also need to be present Year of Birth 4 1880 to current year If Year of Birth is present in the barcode then Day of Birth and Month of Birth also need to be present Age 99 Numeric 9999 9999 Either a period or comma is accepted as a decimal symbol Age Unit 99 A Y YR YRS years MO MOS months WK WKS weeks D DAY DYS days Gender 1 If gender is configured not fixed this mu...

Страница 93: ...monitor is restarted The keyboard locale takes effect immediately after it is submitted 7 11 Selecting national requirements Activate France specific defaults for the ECG HR adjustment range and the reminder beep behavior 1 Log in to Webmin 2 Select Configuration National Requirements 3 In the National Requirement window select the applicable option 4 Select Save The national requirements changes ...

Страница 94: ...w only and cannot be changed To set the host asset number 1 Log in to Webmin 2 Select Configuration Host Asset Settings 3 In the Host Asset Settings window enter the user assigned host asset number up to 32 ASCII characters long in the Change Value to column 4 Select Save The host asset changes take effect immediately 7 14 Changing passwords You can change the passwords for the biomed and clinical...

Страница 95: ...ser s password as required 4 Select Save The change will take effect immediately 7 15 Restarting the patient monitor You can use the Restart module in Webmin to restart the patient monitor after making configuration changes that require a manual restart to come into effect For example after changing network or language settings or adding activation codes for licenses NOTE Loss of monitoring This f...

Страница 96: ...uired for the Remote Service communication to work Otherwise select None If the HTTP proxy server requires user authorization a specific Username and Password is required Otherwise select None These values are determined by the customer Port Username and Password Remote Service Configuration Item Description Comments System ID Identifies the system to the GE back office servers These values are re...

Страница 97: ...he destination patient monitors section 7 17 2 Loading settings 4 Activate the loaded platform and or clinical configuration settings in the destination patient monitors section 7 17 3 Activating settings 5 Some platform settings can t be transferred from one patient monitor to another See section 7 17 3 Activating settings for details Configure these unaffected settings manually in the destinatio...

Страница 98: ...drive to the patient monitor NOTE The loaded settings will remain in an inactive state in the patient monitor until they are purposely activated by the user see section 7 17 3 Activating settings 1 Log in to Webmin 2 Select Configuration Settings 3 In the Sub Modules for Settings menu select Load 4 According to your Webmin access a If you are using Webmin on a service PC you can load the settings ...

Страница 99: ...d platform settings to activate only loaded platform settings NOTE Activation of the cloned clinical or platform settings will not effect the following settings in the target patient monitor Network Unit and Bed Name Host Asset Settings Licenses Remember to configure the unaffected settings manually in the target patient monitor NOTE Check and manually modify the following WLAN settings after acti...

Страница 100: ...ttings activation will start automatically after the next case end patient discharge unless this process is canceled by the user c Wait until the settings activation is completed and the patient monitor has performed an automatic restart d Verify that the settings activation was successful and the patient monitor is running the activated settings 7 17 4 Canceling pending settings activation You ca...

Страница 101: ...e Package window a From the Status drop down list select ENABLED b Enter a valid Activation Code for the new software version 5 Select Save 6 To activate a software package select the desired radio button in the Active column 7 Select Save All license changes take effect after the next patient monitor restart NOTE The software license for the active host software version is shown in the Software P...

Страница 102: ...be uploaded NOTE Do not disconnect the USB storage device until downloading is complete 5 Select Upload to upload the license file 6 Verify that the information populated in the Software Package and Host License tables is accurate All license changes take effect after the next patient monitor restart 7 19 Software management Software installation consists of two main steps software transfer and so...

Страница 103: ...ting clinical and platform settings of the patient monitor are saved and are not affected by the activation of the new patient monitor software version However any new or changed clinical and platform settings in the activated patient monitor software version have their factory default values and may require manual configuration For more information refer to the latest version of the CARESCAPE Mod...

Страница 104: ...ately 4 In Erase Inactive Software After Activation select 5 Select Activate to start the host software activation 6 Confirm activation by selecting Yes in the Host Software window that opens The patient monitor shows a screen saver that informs about the ongoing software activation No to keep the currently active software version as inactive software after the new software version is activated su...

Страница 105: ...or restarts automatically 8 Verify that the software activation is successful and the patient monitor runs the activated software Activating host software after next case end discharge If you want to activate the new host software with a time delay that is after the next case end patient discharge follow this instruction You can initiate the software activation process with a delay during an activ...

Страница 106: ...n is complete and the patient monitor restarts automatically If the patient monitor starts up normally and no error messages appear on the display the activation is successful Activating module software Follow this instruction when you want to activate new sGAS software to a CARESCAPE Respiratory Module activate new PDM software to a CARESCAPE Patient Data Module activate new PiCCO software to a E...

Страница 107: ...lay 1 Log in to Webmin 2 Select Configuration Software Management The Software Management window opens 3 Select Host Software from the Software List and select Next 4 In the Host Software window below Status verify that the Software activation pending is Yes 5 Select Cancel The pending software activation is cancelled and the patient monitor continues running the current software version 7 19 4 Er...

Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...

Страница 109: ...rform the following visual inspection to the installed monitoring system Carefully inspect the patient monitor and the connected peripheral devices for any damage Verify that the patient monitor and the connected peripheral devices are properly mounted with specified mounting solutions Verify that the cables between the patient monitor and the connected peripheral devices are intact properly conne...

Страница 110: ...ate unless a properly wired power outlet is used 8 2 3 Power cord and plug Verify that the power cord being used with the patient monitor is good To do this check the following Inspect the power cord for wear or damage regularly If damage is suspected test for continuity through each conductor of the power cord connector Replace the power cord as necessary with a regulatory approved cord for the c...

Страница 111: ...urrent source with a frequency of 50 or 60 Hz with a no load voltage not exceeding 6 V is passed for at least 5 seconds but not more than 10 seconds through the protective earth terminal or the protective earth pin in the mains plug and each accessible metal part which could become live in case of failure in basic insulation 2 The voltage drop between the parts described is measured and the impeda...

Страница 112: ...his test NOTE The measuring device MD represents the network and voltage measuring instrument and its frequency characters according to IEC 60601 1 1 Configure the safety analyzer as follows NC Polarity NORMAL Neutral CLOSED 2 Power on the device under test 3 Read and record the current leakage indicated on the safety tester 4 Configure the safety analyzer as follows SFC Polarity NORMAL Neutral OP...

Страница 113: ...ngs shall be less than or equal to 1 mA 8 2 6 Enclosure leakage current touch current test This test measures current leakage through the exposed conductive parts on the device under test Perform the test in Normal Condition NC and in two different Single Fault Conditions SFC 1 earth open and 2 one of the supply conductors open at a time Perform the test with normal and reverse polarity NOTE Refer...

Страница 114: ...arity REVERSED Neutral CLOSED Earth GND OPEN 9 Read and record the current leakage indicated on the safety tester 10 Configure the safety analyzer as follows SFC Polarity REVERSED Neutral OPEN Earth GND CLOSED 11 Read and record the current leakage indicated on the safety tester 12 Configure the safety analyzer as follows NC Polarity REVERSED Neutral CLOSED Earth GND CLOSED 13 Read and record the ...

Страница 115: ... to use for each patient connector refer to the service instructions included in the safety test body kit NOTE If not otherwise stated in the table below each test body is connected directly to the specified connector in the patient module Table 1 Patient connectors to be tested with each module Module Patient connector E PSMP E PSMPW ECG SpO2 E PSM E PSMW ECG SpO2 E P E PT P3 P7 E PP P5 E COP E C...

Страница 116: ...g instrument and its frequency characters according to IEC 60601 1 NOTE Perform this test for all the connected parameter modules and patient connectors specified in Table 1 1 Configure the safety analyzer as follows NC Polarity NORMAL Neutral CLOSED Earth GND CLOSED 2 Power on the device under test 3 Read and record the current leakage indicated on the safety tester 4 Configure the safety analyze...

Страница 117: ...14 Power off the device under test 15 Repeat this test for all the connected parameter modules and patient connectors specified in Table 1 Acceptance criteria in Normal Condition NC All readings shall be less than or equal to 10 μA Acceptance criteria in Single Fault Condition SFC earth open or one of the supply conductors open at a time All readings shall be less than or equal to 50 μA 8 2 9 Pati...

Страница 118: ...current leakage indicated on the safety tester 6 Power off the device under test 7 Repeat this test for all the connected parameter modules and patient connectors specified in Table 1 Acceptance criteria All readings shall be less than or equal to 50 μA 8 2 10 Test completion 1 Disconnect the safety analyzer from the power outlet 2 Disconnect the test equipment from the patient monitor 3 Disconnec...

Страница 119: ...at all text is readable and all images are clear 2 Verify that the brightness is good Adjust if necessary Touchscreen control Perform this test both for the integrated main display and for the optional secondary display 1 Verify the operation and the calibration of a touchscreen by touching a corner of an active parameter window Verify that the selected menu is opened 2 Select to close any open me...

Страница 120: ...ctivity is started 3 Select to close any open menu and return to the main display 8 3 6 Mouse Perform this test only if a mouse is connected to the patient monitor 1 Move the mouse until the pointer arrow is over an active parameter window or a main menu item you wish to select and click the left mouse button once to select it 2 Select to close the open menu and return to the main display 8 3 7 Al...

Страница 121: ... is connected to a wireless MC Network or S 5 Network NOTE The wireless network must be properly installed and the patient monitor must be within the wireless coverage area Parser Type Test Procedure Length Delimited or Character Delimited Parser 1 2 3 4 Select Scan from Barcode 5 Scan a known test barcode obtained from the hospital NOTE The barcode data content must be known and in compliance wit...

Страница 122: ...g 8 Type the IP address of the patient monitor you just viewed and select ping Check that the patient monitor replies to the ping command there are no lost packets and the maximum latency is 250 ms NOTE The patient monitor withstands a maximum packet loss of 5 packets per 1 million and maximum latency of 250 ms without performance degradation 9 Select to return to the main display 10 Reconnect the...

Страница 123: ...it 1 Log in to Webmin 2 Select Configuration Remote Service 3 In the Sub Modules for Remote Service menu select Control 4 In the Remote Service Control window under Test Connectivity select Test to test the connectivity to the remote service server 5 Verify that Connection to Enterprise URL and Connection to Tunnel URL both Pass the test If one of the connection tests fail check the failure descri...

Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...

Страница 125: ...Theory of operation 117 2062973 003 9 Theory of operation Figure 3 CARESCAPE Monitor B450 System block diagram The system block diagram shows the functional units of the CARESCAPE Monitor B450 ...

Страница 126: ... and 65 W continuous output power for the DC DC board The AC DC power supply unit has overload and overvoltage protections AC DC Output cable The AC DC output cable provides the 15 7 Vdc to the DC DC board Batteries The patient monitor supports two rechargeable and exchangeable 9 12 3V 3800 mAh lithium ion batteries They are located in the battery compartment One battery must be installed at all t...

Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...

Страница 128: ...er VSYS voltage level depends on the selected power source In battery use the VSYS is between 8 7 12 3V and in AC DC use the VSYS is between 14 9 16 1V The VSYS voltage s supplied to the PDM module and to the voltage converter in the CPU Carrier Board All supply voltages have over voltage and short circuit protection Smart Battery Management The PUIC controller acts as a SMBus host device and take...

Страница 129: ...ight sensor and passes it to the CPU assembly CPU assembly adjusts the brightness of the display according to the ambient light information received from the PUIC controller User interface board The UIC subsystem has the LED drives for three power indicator LEDs and ON Stdby button LED in the keypad The PUIC controller provides an audible power failure alarm to the buzzer in case of sudden system ...

Страница 130: ...ďůĞ YƵĂĚ ƵƉƉůLJ sŽůƚĂŐĞ DŽŶŝƚŽƌ ǁŝƚŚ ĚũƵƐƚĂďůĞ ZĞƐĞƚ dŝŵĞƌ нϱs нϯ ϯs WŽǁĞƌ ƵƉƉůLJ DK K Z ŽŶŶĞĐƚŽƌ D ZDϵ ĐŽŶƚƌŽůůĞƌ Z D ϮϱϲD d E E ůĂƐŚ Ϯ d KKd ůĂƐŚ ϭD d DŽĚƵůĞ ƵƐ Z ϰϴϱ dƌĂŶƐĐĞŝǀĞƌ ZĞĐŽƌĚĞƌ ŝŶƚĞƌĨĂĐĞ ĞĨŝď DĂƌŬĞƌ ŶĂůŽŐ W tĂƚĐŚĚŽŐ ZĞƐĞƚ Z D ϮϱϲD d s h W Ğ sK E ϭϬ ϭϬϬ ϭ ϭϬ ϭϬϬ ϭ s Ͳ d h ƚŽ dŽƵĐŚ ŽŶƚƌŽůůĞƌ h ƚŽ Wh Ϯ нϭϱsDK s z нϭϮs нϯ ϯsͺWD нϱs нϯ ϯs t d ϭϬ ϭϬϬ ǁ Ž ŵĂŐŶĞƚŝĐƐ ϭϬ ϭϬϬ ϭϬ ϭϬϬ D ƚŽ D ϭϬ ϭϬϬ ...

Страница 131: ...xecution and temporary data storage The EMBC section takes care of the following CPU services It provides serial communication for the optional built in recorder It provides RS 485 module bus communication for the E modules and PSM module It provides analog ECG and invasive pressure outputs and a digital marker out signal for a user supplied equipment for example a defibrillator or an intra aortic...

Страница 132: ...k to store system date and time An I2C A D converter and SPI temperature sensor measure the CPU supply voltages and board temperature External Interfaces The CPU carrier board provides the following external interfaces In the rear panel 2 high speed USB connections for the USB input devices for secondary display and for exporting monitoring data to external systems 1 DVI D connector for an externa...

Страница 133: ...n time code execution and temporary data storage The CPU module provides the following interfaces LVDS Low Voltage Differential Signaling interface for the integrated LCD display and SDVO Serial Digital Video Out output to SDVO DVI transmitter in the CPU Carrier Board The Ethernet host controller for the Ethernet switch is in the CPU Module PCIe interface for the Ethernet controller in the CPU car...

Страница 134: ...s used to provide audible alarms The audio signal for the speaker is generated in the CPU carrier board using an audio codec and an audio amplifier Audio feedback controls the operation of the speaker The speaker cable is connected to the E module board 9 4 6 Keypad Buzzer The patient monitor has a user interface board with an on standby button and three power indicator LEDs mains voltage indicato...

Страница 135: ...he CPU carrier board It provides interfaces for PSM and PDM modules The 10 pin PDM connector provides the VSYS PDM supply voltage and the Ethernet communication lines to the PDM module The PSM connector provides the 15 VMod PSM supply voltage and the RS 485 communication lines to the PSM module ...

Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...

Страница 137: ...equipment failure and possible health hazards NOTE The manufacturer does not in any manner assume the responsibility for performing the recommended maintenance schedule unless an Equipment Maintenance Agreement exists The sole responsibility rests with the individuals hospitals or institutions utilizing the device NOTE Refer to Module Frames and Modules Technical Manual for corrective and planned ...

Страница 138: ...orm the following electrical safety tests described in detail in chapter 8 Installation checkout 8 2 1 Test setup 8 2 2 Power outlet 8 2 3 Power cord and plug 8 2 4 Ground earth integrity 8 2 5 Earth leakage current test 8 2 6 Enclosure leakage current touch current test Performed service activity Required checkout procedure Visual inspections section 10 1 Electrical safety test section 10 2 Funct...

Страница 139: ...ion Device Information 3 Verify that the information about the connected module appears in the table named as Acquisition Information E Modules NOTE For some parameter modules the table does not show the actual module information but the information of the individual subassemblies inside the module NOTE You may need to refresh the Webmin screen if you have connected the module to the patient monit...

Страница 140: ...t the recorder starts printing Let the recorder print for approximately 10 seconds and verify the following things from the printout The header line contains the date time and some other applicable status and configuration information The grid is clear The waveforms labels appear in the printout as configured Stop printing by selecting Monitor Setup Printing Waveforms Stop Printing or the related ...

Страница 141: ...ation synchronization connector in the patient monitor 6 Connect the open wires of the other end of the analog output cable to the oscilloscope probe as described in each test Use the figure and table below as a reference for making the connections Brown wire is not connected Configuring the monitor 1 Configure ECG a Select ECG1 to the screen with waveform and adequate priority b In the Parameter ...

Страница 142: ...t the oscilloscope Time and Volts scales c Verify that the ECG analog output waveform shown on the oscilloscope screen corresponds to the ECG1 waveform on the monitor screen 2 Test the Arterial BP output signal NOTE Perform this arterial BP test only if an E PSMPor an E PSMPW module is in use with the patient monitor a Ensure that the simulator s InvBP output channel is configured to 0 mmHg static...

Страница 143: ... Marker out signal a Change the Time scale in the oscilloscope b Verify that the signal shown on oscilloscope screen corresponds to the picture above 10 3 16 Test completion Select Discharge Patient or Reset Case to discard any changes made to the patient monitor configuration during checkout Complete the Appendix B Maintenance check form IP analog output signal Yellow Analog GND Blue or White Pro...

Страница 144: ...rformance Location Position the equipment in a location that does not artificially increase the operating temperature of the batteries Conditioning guideline Condition the battery when a Condition Battery A or Condition Battery B message is shown on the monitor screen The condition cycle recalibrates the electronic fuel gauge 10 4 2 Storage recommendations GE recommends storing the battery outside...

Страница 145: ...blem persists replace the battery NOTE Refer to section 11 6 Battery diagnostics and 11 7 Error messages and codes for more detailed information about the battery status and operating condition 10 4 6 Replacing a battery Replace the battery in the following situations if either Replace battery A or Replace battery B message is displayed This message indicates that the full charge capacity of the u...

Страница 146: ...ormation about returning batteries to GE contact your authorized GE Service representative or contact GE Equipment Services at 1 800 437 1171 Canada Contact the approved battery stewardship program in your province for information on recycling your batteries Other countries Recycle batteries through your local regional or national collective scheme in accordance with your local or national regulat...

Страница 147: ... that There is no physical damage All peripheral devices are connected properly The patient monitor and the connected peripheral devices are properly powered Also verify that the problem is not caused because of 1 Incompatibility issue See the patient monitor s supplemental information manual for the list of compatible devices 2 Incorrect platform or clinical configuration Refer to the chapter 7 C...

Страница 148: ...umber and version Host serial number Host asset number MC Network IP address IX Network IP address MAC address S 5 Network virtual ID CPU hardware version EMBC software version PUIC software version and DC DC hardware version Host Hardware Information Status information for the number of installed batteries recorder and WLAN options PDM License Information PDM license option status and number of l...

Страница 149: ... Remote service status System ID Serial number Enterprises URL Enterprises tunnel URL and Protocol Language Clinical user interface language National Requirement Setting for country specific alarms None or France Network Active configuration information including MAC address MC Network type IP address Netmask Gateway Destination IP address Destination netmask and PHY configuration and IX Network t...

Страница 150: ...and feature code Active Software Package Current software package in use status and feature code Host License Information Each host license name its current status enabled disabled or trial feature code and the expiration date for a trial license Acquisition Information PDM Active software version Main board revision DAS board revision Serial number Asset number MAC address IP address Power freque...

Страница 151: ...if it cannot be measured To access hardware statistics 1 Log in to Webmin 2 Select Diagnostics Hardware Stats 3 Scroll down the page to view the following device information The controlled parameters are measured with A D converters and temperature sensors in the specified subsystem The measured values are then communicated over the I2C bus to the power management controller in the DC DC board for...

Страница 152: ...tery VSYS voltage mV 12V voltage mV 15V voltage mV These voltages are generated by the voltage converters in the DC DC board These voltages supply power to the following subsystems VSYS 9V 16V is supplied for the PDM module or ePort 12V is supplied for the CPU module display backlight and the optional recorder 15V is supplied for the E modules and PSM module when 15VMOD circuit breaker is enabled ...

Страница 153: ...onverter in the CPU carrier board VSYS mV This voltage is generated by the voltage converter in the DC DC board It is measured by the A D converter in the CPU carrier board VSYS_EPORT mV The VSYS_EPORT voltage is supplied for a PDM module that is connected to the ePort connector DB9 in the rear of the patient monitor It is measured by the A D converter in the CPU carrier board The value is shown a...

Страница 154: ...n active network The patient monitor withstands a maximum packet loss of 5 packets per 1 million and maximum latency of 250 ms without performance degradation 11 3 3 WLAN diagnostics Use this Webmin feature for troubleshooting WLAN related problems This Webmin module provides information about the WLAN driver WLAN status and the detected access points 1 Log in to Webmin 2 Select Diagnostics WLAN 3...

Страница 155: ...egory Voice Video Best Effort Background The DSCP settings for network traffic types Realtime Clinical Traffic Non Realtime Clinical Traffic Non Realtime Non Clinical Traffic SSID i e Service Set Identifier network name Security status Authentication method i e open Confidentiality method i e None WEP 64 WEP 128 WPA PSK TKIP WPA2 PSK AES CCMP Used Key Index if WEP 64 or WEP 128 is used RF readings...

Страница 156: ...n terms of dBm The minimum RSSI should be 60dBm 5dB 54Mbps or greater to maintain application performance Operating channel number MAC Address Log Contents Association Authentication state Description of the state Disconnected The WLAN client radio is not authenticated or associated to any network infrastructure Authenticating The network infrastructure is authenticating the WLAN client radio Auth...

Страница 157: ...tion occurred Module software activation information including the module type and serial number the type and version of the activated module software and the date and time the module software activation occurred Settings transfer history information including the type of the transferred settings the origin and destination and the date and time the settings transfer occurred Log file transfer hist...

Страница 158: ... log file to GE Service for further investigation To view log files 1 Log in to Webmin 2 Select Diagnostics View Logfiles PDM log All PDM errors and messages System log OS events and errors including operating system related information such as clinical application startup and recovery information power on self test results etc PUIC PMC WLAN and battery events and errors InSite with ExC events and...

Страница 159: ...troubleshooting LEDs in the DC DC board are visible through the ventilation holes in the rear of the patient monitor see picture below LED Description On standby red H6 The red on standby LED should be lit momentarily when the on standby button is pressed A lit LED normally indicates that the DC DC power supply board recognizes that the user has pressed the on standby button and should turn the pa...

Страница 160: ...ne of the problems Main fuses are blown AC DC cable is disconnected from DC DC board Faulty AC DC power supply unit especially if the patient monitor operates normally when it is battery powered Faulty DC DC board especially if the patient monitor does not operate neither on AC mains nor battery 12 V green H3 These secondary voltages are created by the DC DC board and they should be lit when the p...

Страница 161: ...ation failure error condition is on Failure Battery error condition is on Discharging Patient monitor is operating on battery power Charging Patient monitor is connected to AC mains and charging the battery Full Patient monitor is connected to AC mains and battery is fully charged Time to full hh mm hh mm Estimated time to charge the battery to full capacity N A Battery is not connected to the pat...

Страница 162: ... capacity of a new battery Cycle count The total count of charging and discharging cycles of the battery Voltage V Battery voltage Current mA Battery current Positive when charging negative when discharging Temperature C or F Battery temperature Message Possible causes Possible solutions License expired A trial license has expired Enable the license with a new activation code if needed Configurati...

Страница 163: ...ts and other electrical faults 6 If possible log in to Webmin and select Diagnostics Hardware Stats to identify the root cause for the error message If temperature is too low or high If the patient monitor has been transported or stored outside the operating temperature range allow it to stabilize back to operating temperature range before applying power Service Monitor Error Code 0xHOST1002 One o...

Страница 164: ...e end next discharge If required the service user may cancel the delayed setting activation See section 7 17 4 Canceling pending settings activation Software activation after next case end Software activation after next discharge Service user has initiated a delayed software activation that will automatically take place after next case end next discharge If required the service user may cancel the...

Страница 165: ...ry B Battery needs conditioning Condition the monitor battery See section 10 4 5 Conditioning a battery Battery A temperature high Battery B temperature high Battery temperature error due to faulty battery or battery management error Replace battery with a new one If that does not help replace the DC DC board Speaker failure 1 Speaker cable loose 1 Connect the speaker cable 2 Speaker failure 3 Rep...

Страница 166: ...e power outlet being used Refer to the patient monitor s supplemental information manual for power requirements Check the power outlet being used see section 8 2 2 The cable between the AC DC power supply unit and the DC DC board is loose or faulty Check that the cable is intact and properly connected to the AC DC power supply unit and the DC DC board The AC DC power supply is faulty Replace the A...

Страница 167: ...the cable if loose Replace the user interface cable if the cable is faulty The user interface cable is loose or faulty Connect or replace the user interface cable The user interface board is faulty Replace the user interface board DC DC Board failure Replace the DC DC board Unable to turn on the patient monitor The patient monitor starts but the primary display remains black i e the backlight does...

Страница 168: ... appear to the screen 1 uDOM is missing or loose 2 uDOM software is corrupted 3 uDOM has incompatible software 1 Ensure that the uDOM is properly connected and aligned to the connector in the CPU carrier board 2 Replace uDOM Contact GE service for support 3 Ensure that the attached uDOM has correct and compatible software Contact GE service for support Webmin login screen appears after the startup...

Страница 169: ...m light illuminates during the power on self test only if it is a cold start more than 15 minutes from the previous start up Problem Possible cause Recommended action Wrong character is displayed when a key is pressed on keyboard The keyboard locale is not configured correctly Configure the keyboard locale correctly see 7 10 Wrong character is displayed when a barcode is read The keyboard locale i...

Страница 170: ...nt monitor is not connected to network CPU timekeeper battery empty Replace CPU timekeeper battery Time not configured properly Configure date and time see 7 3 System time is incorrect when patient monitor is connected to MC Network Network device time synchronization error When adding a new device to the CARESCAPE Network the existing devices on the CARESCAPE Network will synchronize to the new d...

Страница 171: ...figuration Licenses If you have printed license information select Software Package and Host Licensing If you have a license file select Upload License The wrong software application is displayed on the patient monitor The wrong software application is activated for the device 1 To view the software package that is currently activated Log in to Webmin Configuration Licenses Host Licensing 2 Make s...

Страница 172: ...most likely in the patient monitor Continue troubleshooting the problem according to the related troubleshooting chart below No The suspect module is most likely faulty Refer to Module Frames and Modules Technical Manual for troubleshooting instructions Problem Possible cause Recommended action Recorder does not work Graph location is not configured correctly Check the configuration Monitor Setup ...

Страница 173: ... is most likely in the CPU carrier board Replace the CPU assembly Possible cause Recommended action Pole mount cable loose or faulty if the PSM module is mounted to a pole mount Check that the Module Bus Adapter for PSM and the connection cable is intact and properly connected to the PSM module mounted to the IV pole and to the PSM connector in the patient monitor DC DC board failure The 15VMOD su...

Страница 174: ... board failure The VSYS supply voltage for the PDM module is generated in the DC DC board Check from the Webmin Hardware Statistics that the supply voltages are within specification VSYS voltage mV VSYS_EPORT mV VSYS_PDM mV If PDM is connected to PDM dock DC DC board should identify PDM and enable the VSYS_PDM circuit breaker on CPU carrier board to connect VSYS voltage to VSYS_PDM If PDM is conne...

Страница 175: ...Troubleshooting 167 2062973 003 ...

Страница 176: ...DQ RX VHOHFW DQRWKHU SDWLHQW PRQLWRU WR YLHZ IURP WKLV SDWLHQW PRQLWRU HV QR 1HWZRUN SUREOHP URDGFDVW WUDIILF LV JRLQJ EHWZHHQ WKH SDWLHQW PRQLWRUV EXW QRW WR WKH 3UR FHQWHUV 7KH 3UR FHQWHUV PD QRW EH RQ WKH VDPH QHWZRUN DV WKH SDWLHQW PRQLWRUV 3OHDVH YHULI FRQILJXUDWLRQ VHWWLQJV XVLQJ WKH LUHOHVV 1 1HWZRUN RQILJXUDWLRQ XLGH DQG FRQVXOWLQJ WKH KRVSLWDO V 7 GHSDUWPHQW RQILJXUDWLRQ LVVXH 6HOHFW WKH ...

Страница 177: ...N SUREOHP 3OHDVH YHULI FRQILJXUDWLRQ VHWWLQJV ZLWK WKH KRVSLWDO V 7 GHSDUWPHQW HV 8VHU UURU 3OXJ LQWR FRUUHFW MDFNV QR RQILJXUDWLRQ LVVXH RUUHFW 3 DGGUHVVLQJ LVVXHV HV 1R WUDIILF LV JHWWLQJ WKURXJK 3RWHQWLDO QHWZRUN FRQILJXUDWLRQ LVVXH V WKH 66 WLHG WR WKH FRUUHFW 9 1 3 FRQILJ V WKH 9 1 EULGJHG WR WKH 5 6 3 1HWZRUN V WKH 3UR FHQWHU LQ WKH FRUUHFW 5 6 3 1HWZRUN V WKHUH D ILUHZDOO RU FFHVV RQWURO LV...

Страница 178: ...all bedside monitors affected The problem is related to the iCentral or to the cabling between the iCentral and the hub switch Proceed to the iCentral troubleshooting Check the cabling between the iCentral and the hub switch Bedside monitor doesn t register to the Monitor Network only one patient monitor or some of the bedside monitors affected The problem is related to the bedside monitor or to t...

Страница 179: ...ess network infrastructure IEEE 802 11b SSID mismatch between patient monitor and access point Enter correct SSID see section 7 2 4 Security settings mismatch between patient monitor and access point Incorrect or incompatible confidentiality method Incorrect Pass Key Incorrect key index Contact the hospital IT department for correct security settings and then update the patient monitor configurati...

Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...

Страница 181: ...ndling semiconductors or assemblies containing semiconductors A grounded antistatic wristband or heel strap should be worn at all times while handling or repairing assemblies containing semiconductors Use properly grounded test equipment Use a static free work surface while handling or working on assemblies containing semiconductors Do not remove semiconductors or assemblies containing semiconduct...

Страница 182: ...ge the existing thread patterns However when using battery operated tools ensure that it is equipped with torque limiter and the torque is properly adjusted 12 1 3 Required tools NOTE Use torque wrench and torque screwdriver to comply with the given torques WARNING Due to possible high voltage present use an insulated screwdriver at all times 12 1 4 Before disassembly WARNING PATIENT MONITORING IN...

Страница 183: ...monitor power cord first from the wall outlet and then from the monitor 3 Remove the patient monitor batteries 4 Disconnect all external cables connected to the patient monitor 5 Disconnect all parameter modules 6 Detach the patient monitor from the mount if installed NOTE If you disconnect the patient monitor from the AC power source and remove the battery prior to turning the monitor off the mon...

Страница 184: ...ƐĞŵďůLJ ĨƌŽŵ ƚŚĞ ƌĞĂƌ ƵŶŝƚ ĂƐƐĞŵďůLJ ZĞĂƌ ƵŶŝƚ ĂƐƐĞŵďůLJ ƌŽŶƚ ƵŶŝƚ ĂŶĚ ŵŝĚͲ ĨƌĂŵĞ ĂƐƐĞŵďůLJ WŽǁĞƌ ƵƉƉůLJ hŶŝƚ Zh Wh ĂƐƐĞŵďůLJ Zh ĞƚĂĐŚŝŶŐ ĨƌŽŶƚ ƵŶŝƚ ĂƐƐĞŵďůLJ ĨƌŽŵ ŵŝĚͲĨƌĂŵĞ ĂƐƐĞŵďůLJ ĚŝƐƉůĂLJ ƵŶŝƚ Zh hƐĞƌ ŝŶƚĞƌĨĂĐĞ ďŽĂƌĚ Θ ďƵnjnjĞƌ Zh ƌŽŶƚ ƵŶŝƚ ǁ ƚŽƵĐŚƐĐƌĞĞŶ ƐĞŶƐŽƌ Zh ƉĞĂŬĞƌ ƵŶŝƚ Zh ͲDŽĚƵůĞ ďŽĂƌĚ Zh ďŽĂƌĚ Zh DŽĚƵůĞ ƌĂŝů ƵŶŝƚ Zh t E ĂƐƐĞŵďůLJ Zh ĞƚĂĐŚŝŶŐ ƵƐĞƌ ŝŶƚĞƌĨĂĐĞ ďŽĂƌĚ ĂŶĚ ďƵnjnjĞƌ ĨƌŽŵ ƚŚĞ ĨƌŽŶƚ ƵŶŝƚ ĂƐƐ...

Страница 185: ... the recorder by pressing with a flat blade screwdriver 3 Pull the recorder out of the monitor first a little by pulling from the recorder door and then all way out by pulling from the recorder body NOTE Be careful not to damage the recorder door or the grounding plate when pulling out the recorder FRU Recorder unit B450 Reassemble in reverse order Push the recorder unit all the way into its slot ...

Страница 186: ...nitor inside the AC inlet 1 Release the snap in the fuse holder of the AC inlet by pressing it with a flat blade screwdriver Pull the fuse holder out of the AC inlet 2 Replace the fuses with new ones Use only fuses with the same rating Reassemble in reverse order Push the fuse holder all the way in until it snaps to its place ...

Страница 187: ...k device time synchronization issues remember to readjust the time and date before you connect the monitor to network For details see Setting time and date 1 Remove the screw T10 holding the service cover Torque 0 6 Nm 2 Release the snap that holds the service cover using a flat blade screwdriver ...

Страница 188: ...rrier board with a flat blade screwdriver NOTE Dispose the battery according to local state or country laws Reassemble in reverse order To avoid network device time synchronization issues remember to readjust the time and date before you connect the monitor to network For details see 7 3 Setting time and date ...

Страница 189: ...ry door assembly to the patient monitor using a flat blade screwdriver 3 Pull the battery door assembly out of the patient monitor FRU Battery door unit B450 Reassemble in reverse order Align the battery door plate to its guide rails Push the battery door assembly all the way into the monitor frame until it snaps to its place Close the battery door ...

Страница 190: ...rque during the reassembly if you replaced the CPU assembly use 1 2 Nm if you didn t replace the CPU assembly use 0 8 Nm 2 Pull out the module rail unit FRU Module rail unit B450 Reassemble in reverse order Ensure that the module rail unit connector is correctly aligned to the connector in the CPU board WARNING Use only GE specified screws and torque when replacing the module rail unit ...

Страница 191: ...1 Use a flat blade screwdriver to release the snap that fastens the recorder cover plate to the monitor frame Detach the recorder cover plate NOTE If there is a recorder installed you do not have to remove it 2 Remove the screw T10 holding the service cover Torque 0 6 Nm 3 Release the snap that holds the service cover using a flat blade screwdriver 4 Pull out the service cover ...

Страница 192: ...lens by sliding a flat blade screw driver under the lens and carefully lift the lens up using the screw driver 6 Remove the three screws T10 in the bottom of the patient monitor Torque 0 6 Nm 7 Remove the two screws T10 in the top of the patient monitor Torque 1 2 Nm 8 Remove the screw T10 next to the handle Torque 1 2 Nm ...

Страница 193: ...sembly 185 2062973 003 9 Pull out the front unit and mid frame assembly from the rear unit assembly a little so that you have access to the AC DC and display cables 10 Detach the AC DC cable from the DC DC board connector ...

Страница 194: ...l out the front unit and mid frame assembly all way out from the rear unit assembly Reassemble in reverse order When reassembling ensure that the connector in the E module board is properly aligned with the connector in the CPU carrier board Remember to reconnect the AC DC cable and the display cable ...

Страница 195: ...t from the other end You may need pliers to squeeze the nylon support post a little to detach the uDOM NOTE uDOM disk on module is the permanent memory of the patient monitor The software licenses and clinical and service configurations are stored to the uDOM Reassemble in reverse order uDOM replacement Contact GE Service if you doubt that the uDOM is defective and needs to be replaced Some of the...

Страница 196: ...right on the table Support the CPU board and AC DC power supply unit with your hand so that they do not drop when you perform the following steps 1 Remove the four screws T10 that mount the module rail unit to the rear unit assembly NOTE Notice the torque during the reassembly if you replaced the CPU assembly use 1 2 Nm if you didn t replace the CPU assembly use 0 8 Nm 2 Pull out the module rail u...

Страница 197: ... to fasten the grounding screw properly This is important to ensure that the CPU assembly is properly connected to the protective earth 5 Release the latch holding the CPU assembly and pull the assembly out from the rear unit 6 Carefully detach the uDOM from the CPU assembly The uDOM is connected to the CPU assembly with a connector from one and with a nylon support post from the other end You may...

Страница 198: ...order Note the following Step 6 Remember to first reattach the original uDOM to the CPU assembly Step 5 When reassembling the CPU assembly make sure to first place the heat sink under the plastic lip of the rear unit assembly Then press the other edge of the heat sink under the latch ...

Страница 199: ... the CPU assembly and the AC DC power supply with your hands while you attach and tighten the screws NOTE Notice the torque during the reassembly if you replaced the CPU assembly use 1 2 Nm if you didn t replace the CPU assembly use 0 8 Nm WARNING Use only GE specified screws and torque when replacing the module rail unit NOTE After replacing the CPU Assembly during the first start up the patient ...

Страница 200: ...eep the rear unit assembly on the table the device label side facing down 1 Press the latch and pull out the AC DC power supply unit from the rear unit NOTE Be careful not to damage the fragile WLAN antenna boards or antenna cables Reassemble in reverse order Ensure that the AC DC power supply unit is properly aligned to the rear unit assembly Make sure that the primary conductors are not squeezed...

Страница 201: ...bly from the rear unit assembly 1 Keep the rear unit assembly on the table the device label side facing down Be careful not to damage the fragile WLAN antenna boards and antenna cables 2 Remove the two screws T6 that hold the WLAN radio card to the CPU assembly Torque 0 3 Nm Detach the WLAN radio card carefully from its connector in the CPU carrier board ...

Страница 202: ...ully from their holders on the inner roof of the rear unit FRU WLAN assembly B450 Reassemble in reverse order Be careful not to bend the antenna boards or damage the antenna cards Ensure that the antenna boards and antenna cables are properly mounted to their holders on the inner roof of the rear unit ...

Страница 203: ...ver plate at the bottom of the front panel 2 Remove the two screws T10 holding the front unit assembly to the mid frame assembly Keep the front unit and mid frame assembly on the table the display facing down Protect the touchscreen sensor from any scratches and dust 3 Detach user interface and touchscreen cables from the DC DC board 4 Release the four snaps two on each side that mount the front u...

Страница 204: ...e you lift the mid frame assembly from the front unit assembly The LCD display should remain attached to the display holder of the mid frame unit with the two rubber fasteners of the display gasket Guide the disconnected cables through the openings in the unit Front unit assembly ...

Страница 205: ...hing the front unit assembly back to the mid frame assembly Guide the touchscreen and user interface cables through the openings in the mid frame assembly Connect the cables to the connectors in the DC DC board NOTE You may find it easier to first connect the user interface cable to the DC DC board and guide it towards the user interface board through the opening in the mid frame assembly ...

Страница 206: ...the rear unit assembly b 12 2 11 Detaching the front unit assembly from the mid frame assembly 1 Detach the LCD backlight cable from the DC DC board using a screw driver 2 Guide the display cable through the openings in the mid frame unit 3 Detach the two rubber fasteners that hold the LCD display attached to the plastic display holder of the mid frame unit ...

Страница 207: ...he LCD display with its gasket from the display holder in the mid frame unit FRU LCD Display unit B450 Reassemble in reverse order 1 Guide the display and LCD backlight cables back through the openings in the mid frame assembly 2 Check that the display gasket is properly aligned around the LCD display Then carefully align the LCD display with the gasket to the display holder in the mid frame assem...

Страница 208: ...oard and guide it through the openings in the mid frame unit 2 Lift the speaker up using a screw driver and detach the speaker FRU Speaker Unit B450 Reassemble in reverse order While reattaching make sure to guide the speaker cable correctly through the cable clamp and the guiding notch ...

Страница 209: ...nt unit assembly from the mid frame assembly Keep the front unit assembly on the table the touchscreen sensor facing down Protect the touchscreen sensor from any scratches and dust 1 Remove two screws T10 that mount the user interface board to the front unit Torque 0 6 Nm 2 Detach the buzzer cable from the user interface board 3 Lift the buzzer up The front unit with the touchscreen sensor is a FR...

Страница 210: ...APE Monitor B450 202 2062973 003 FRU User Interface Board Buzzer B450 FRU Front unit with touchscreen sensor B450 Reassemble in reverse order Note Make sure to guide the buzzer cable through the guiding notch ...

Страница 211: ...ly 1 Remove the screw cover plate at the bottom of the front panel 2 Remove the two screws T10 holding the front unit assembly to the mid frame assembly 3 Detach the alarm light board cable from the DC DC board 4 Release the four snaps two on each side that mount the front unit assembly to the mid frame assembly to be able to detach the alarm light board and detach the alarm light board ...

Страница 212: ...CARESCAPE Monitor B450 204 2062973 003 5 Release the snap that holds the alarm light board FRU Alarm light unit B450 Reassemble in reverse order ...

Страница 213: ... assembly from the rear unit assembly 1 Detach the following cables the touchscreen cable and the user interface cable from the DC DC board the speaker cable from the E module board 2 Detach the following cables from the DC DC board Alarm light board cable LCD backlight cable 3 Remove the screw T10 that mounts the DC DC board to the mid frame assembly Torque 0 6 Nm ...

Страница 214: ...ade screw driver to release the DC DC board and the E module board assembly from the mid frame assembly 6 Slide the DC DC board and E Module board assembly out from the mid frame assembly NOTE When reassembling make sure that the connector between the battery board and the DC DC board is properly connected 7 Detach the E module board from the DC DC board ...

Страница 215: ...verse order NOTE After replacing the DC DC Board during the first start up the patient monitor will automatically check the PUIC software version in the replaced DC DC board and update the software if necessary Wait for 5 minutes to see if the software update is initiated and do not interrupt the process ...

Страница 216: ...DC board and E Module board assembly Steps 1 to 3 only for monitors with a recorder 1 Press the two snaps on each side of the recorder with a flat blade screwdriver NOTE If the monitor is not equipped with a recorder there is a cover plate which you do not have to remove 2 Press the grounding plate to release the recorder unit from the mid frame assembly 3 Pull the recorder out of the mid frame as...

Страница 217: ...e mid frame assembly The tip of the screwdriver seen through the open battery slot 5 Pull the battery board out from the mid frame unit assembly Reassemble in reverse order Ensure that the battery board is correctly aligned to its guide rails and DC DC board battery board connector is properly connected FRU Battery board B450 ...

Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...

Страница 219: ...e rail unit assembly with the 4 mounting screws 5 M1233489 FRU Battery Door Unit B450 Includes a complete battery door assembly 6 M1233490 FRU LCD Display Unit B450 Includes the LCD display with LED backlights display cable LED backlight cable and the gasket around the display 7 2093301 001 FRU Front Unit w Touchscreen Sensor B450 Includes the complete plastic front bezel subassembly with the touc...

Страница 220: ... Board B450 Includes the battery board 16 M1233498 FRU E Module Board B450 Includes the E Module board with the mounting screws 17 M1233499 FRU WLAN assembly B450 Includes the complete WLAN assembly with WLAN radio card and 2 mounting screws antenna cables and antenna boards 18 2086113 001 FRU Plastics Kit B450 Includes the following plastic parts service cover middle frame power cable w lock spri...

Страница 221: ...cables between the patient monitor and the connected devices are intact properly connected and secured to the right connectors The modules are properly connected and locked Electrical safety tests Date of manufacture of the device Is there less than 12 months since the device was manufactured Yes Yes you do not have to perform the electrical safety tests To continue the installation check mark the...

Страница 222: ...dition NC Polarity NORMAL 10 µA Single Fault Condition SFC Ground closed normal 50 µA Ground open normal 50 µA Ground closed reverse 50 µA Ground open reverse 50 µA Normal Condition NC Polarity Reserved 10 µA Patient sink leakage current tests Polarity NORMAL 50 µA Polarity Reserved 50 µA Module type S N Observed result Acceptance criteria PASS N A FAIL Patient source leakage current tests using a...

Страница 223: ...d menu opens and selected activity starts Keypad and remote Trim Knob Selected menu opens and selected activity starts Mouse Selected menu opens and selected activity starts Alphanumeric keyboard Keyboard language configuration is correct Barcode reader Data is correctly populated MC Network and S 5 Network Network symbol and other monitor s patient data is correctly shown on the screen Wireless L...

Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...

Страница 225: ...eripheral devices are undamaged The monitor and the connected peripheral devices are properly mounted with specified mounting solutions The cables between the patient monitor and the connected devices are intact properly connected and secured to the right connectors The modules are properly connected and locked Electrical safety tests Power outlet is correctly wired Power cord and plug are undamag...

Страница 226: ...ected activity starts Keypad and remote Trim Knob Selected menu opens and selected activity starts Mouse Selected menu opens and selected activity starts Alphanumeric keyboard Keyboard language configuration is correct Barcode reader Data is correctly populated MC Network and S 5 Network Network symbol and other monitor s patient data is correctly shown on the screen Wireless LAN WLAN network symb...

Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...

Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...

Страница 229: ...nformation Center with full disclosure license NOTE CIC with full disclosure license is optional It is needed for documentation and reporting purposes only It enables you to view afterwards potential network connectivity issues between the transport monitor and the wireless MC Network and to print reports about waveform loss situations Hardwired CARESCAPE modular monitor the stationary monitor A C...

Страница 230: ...t the following aspects when preparing a test plan Identify areas with known or obvious low signal strength Identify areas with known sources of radio frequency interference causing high noise floor and or poor signal to noise ratio Identify the special characteristics in the building layout floors wings patient rooms and construction material used Identify the time and areas of congestion with hi...

Страница 231: ...reen with adequate priority 6 Select the Setup tab in the ECG menu and configure ECG1 lead II ECG2 lead V1 ECG3 lead aVL 7 Start a patient case admit a patient in Data Pages Admit Discharge menu Wireless CARESCAPE Monitor B450 the transport monitor NOTE The PDM PSM module ECG cables and patient simulator for the wireless CARESCAPE Monitor B450 are needed only if a CIC with full disclosure license ...

Страница 232: ...est plan Contact the nursing staff to ensure access to the needed areas before starting the test 1 Move the roll cart to the starting point of the planned test route 2 Stop at each test point and perform the following tasks a On the transport monitor Verify that ECG waveforms from the remote stationary patient monitor display in the bed to bed view without any losses b On the service PC Refresh We...

Страница 233: ... waveform loss is normal during roaming If longer or repeating waveform losses occur between test points make this an extra test point and report it the same way as observations in step 2 Summary and Reporting Include the following documents to the test results 1 Print the full disclosure reports from the CIC about the observed waveform loss situations 2 Print the snapshots of the WLAN diagnostics...

Страница 234: ...CARESCAPE Monitor B450 C 6 6 2062973 003 Test Form Observations Test summary and recommended actions Test point Time hh mm ss RSSI dBm Transmit rate Mbps Observations Date time of testing Signature ...

Страница 235: ......

Страница 236: ...s Information Technologies Inc 8200 West Tower Avenue Milwaukee WI 53223 USA Tel 1 414 355 5000 1 800 558 5120 US only Fax 1 414 355 3790 Asia Headquarters GE Medical Systems Information Technologies Asia GE China Co Ltd No 1 Huatuo Road Zhangjiang Hi tech Park Puding Shanghai P R China 201203 Tel 86 21 5257 4650 Fax 86 21 5208 2008 www gehealthcare com ...