Maintenance and checkout
133
2062973-003
−
Analog output cable (2000633-001)
NOTE: You can alternatively use the GE defib sync tester, 2040582-001, together with the
Analog output cable, 2000633-001 and a multiparameter simulator to perform this test. Follow
the instructions included with the tester. See analog output cable wire colors and related
signals from the table below.
Connecting cables
1.
Ensure that the module is connected to the patient monitor.
2.
Connect the 5-lead ECG trunk cable to the green ECG connector in the module.
3.
Connect the 5-leadwire sets to the trunk cable and to the simulator.
4.
Connect the invasive pressure adapter cable to the simulator and the other end to the
module.
5.
Connect the analog output cable to the defibrillation synchronization connector in the
patient monitor.
6.
Connect the open wires of the other end of the analog output cable to the oscilloscope
probe as described in each test. Use the figure and table below as a reference for making
the connections.
*) Brown wire is not connected.
Configuring the monitor
1.
Configure ECG:
a.
Select
ECG1
to the screen with waveform and adequate priority.
b.
In the
Parameter Setup > ECG > Setup
, select
ECG1 Lead: II.
2.
Configure invasive pressure:
a.
Select
P1
to the screen with waveform and adequate priority.
b.
In the
Parameter Setup > Invasive Pressures > P1
, select:
−
Label: P1.
Defibrillation synchronization connector
Related wire color in the open
end of the Analog output cable
*)
Pin number
Signal
1
Digital defibrillator synchronization
marker out signal
Black
2
Digital defibrillator synchronization
marker in signal
Green
3
Common GND
Red
4
Analog GND
Blue
5
Analog GND
White
6
IP analog output
Yellow
7
ECG analog output
Grey
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......