Installation checkout
111
2062973-003
monitor batteries and disconnecting the power cord from the wall outlet for a moment before
turning the monitor on.
Verify that the patient monitor starts up normally:
−
The yellow, red and blue alarm lights are lit momentarily.
−
The speaker gives an audible beep.
−
The normal monitoring screen appears and there are no error messages on the
screen.
NOTE: Refer to section
to see the procedure for battery conditioning and
battery replacement, if you receive a
Condition Battery A or Condition Battery B
or a
Battery
A failure or Battery B failure
message.
NOTE: Before taking the patient monitor into use for the first time, the battery should be fully
charged. Keep the monitor connected to the mains until the battery is fully charged.
8.3.2 Display
Picture quality
Perform this test both for the integrated main display and for the optional secondary display.
1.
Verify that all text is readable and all images are clear.
2.
Verify that the brightness is good. Adjust if necessary.
Touchscreen control
Perform this test both for the integrated main display and for the optional secondary display.
1.
Verify the operation and the calibration of a touchscreen by touching a corner of an
active parameter window. Verify that the selected menu is opened.
2.
Select
to close any open menu and return to the main display.
8.3.3 Device Information
1.
Log in to Webmin.
2.
Select
Information >
Device Information
.
3.
Verify that the device information is correct:
−
The connected parameter modules are identified.
−
The connected CARESCAPE Network ID interfaces are identified.
−
The connected USB input devices are identified.
4.
Stay connected to Webmin.
8.3.4 Configuration Information
1.
Select
Information >
Configuration Information
.
2.
Verify that the patient monitor is correctly configured:
−
The host Information is correct.
−
The active software package is correct.
−
The correct host licenses are enabled.
−
The correct PDM licenses are enabled.
−
Patient ID prefix is correctly configured.
−
Unit and Bed name are correctly configured.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......