CARESCAPE Monitor B450
156
2062973-003
External alarm
light disconnect.
Check USB
connection.
1. The USB cable between the
patient monitor and the
secondary display is
disconnected when the
patient is admitted.
1. Reconnect the USB cable.
2. Secondary display turned to
standby when the patient is
admitted.
2. Turn on the secondary display.
NOTE: Select
Audio Pause
to acknowledge the message, if the USB
cable is disconnected or the secondary display turned off on purpose.
3. User interface board
communication failure.
3. Update PUIC (Power Management
and User Interface Controller) software
(host software) or replace user
interface board.
Power
management
failure
Power Management
Controller (communication)
failure.
Update PUIC software (host software)
or replace the DC/DC board.
Setting
activation after
next case end./
Setting
activation after
next discharge.
Service user has initiated a
delayed settings activation
that will automatically take
place after next case end. /
next discharge.
If required, the service user may cancel
the delayed setting activation. See
section
Software
activation after
next case end./
Software
activation after
next discharge.
Service user has initiated a
delayed software activation
that will automatically take
place after next case end /
next discharge.
If required, the service user may cancel
the delayed setting activation. See
section
7.19.3. Canceling pending host
Service Monitor
Activation Failed
Software activation failed.
Reactivate software. If that does not
help, reinstall the software and then
reactivate it.
Setting activation failed.
Reactivate settings.
Module
voltage low
•
One of the supply voltages
for the acquisition
modules is out of
specification.
•
An analog output voltage
is out of specification.
Log in to Webmin and select
Diagnostics > Hardware Stats
to
diagnose the problem. If there is a
faulty board, replace it.
Application
error:
Webmin
The operating system informs
that the Webmin local
browser was terminated
abnormally.
Select
Audio Pause
to acknowledge
the message.
Message
Possible causes
Possible solutions
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......