Installation checkout
115
2062973-003
8.3.12 iPanel connection
Perform the following test only if the Citrix / iPanel is configured and in use.
1.
Select
Data & Pages > iPanel
and verify that the initial program (configured in Webmin) is
launched correctly.
2.
Select
to exit the Citrix thin client.
8.3.13 Insite with EXC
Perform the following test only if remote service is configured and enabled, and you did not test
the connectivity already while you enabled it.
1.
Log in to Webmin
2.
Select
Configuration > Remote Service
.
3.
In the
Sub-Modules for Remote Service
menu, select
Control
.
4.
In the
Remote Service Control
window under
Test Connectivity
, select
Test
to test the
connectivity to the remote service server.
5.
Verify that
Connection to Enterprise URL
and
Connection to Tunnel URL
both "Pass" the
test.
If one of the connection tests fail, check the failure description given and fix the connectivity
issue. Re-test the connectivity.
8.3.14 Test completion
Discharge Patient
or
Reset Case
to discard any changes made to the patient monitor
configuration during checkout.
•
Complete the Appendix
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......