Maintenance and checkout
137
2062973-003
10.4.5 Conditioning a battery
Condition the battery when a
Condition Battery A
or
Condition Battery B
message is shown
on the monitor screen. The condition cycle recalibrates the electronic fuel gauge.
Condition the battery by fully discharging and recharging the battery twice according to the
following procedure:
NOTE: The patient monitor must be in a discharged state during battery conditioning.
Disconnect any acquisition modules from the patient monitor if connected.
1.
Turn on the patient monitor. Disconnect the power cord from the wall outlet. Leave the
patient monitor on until the battery is fully discharged and the patient monitor turns off
automatically.
NOTE: Ignore the
Monitor Battery Low
and
Monitor Battery Empty!
messages when
discharging the battery.
2.
Reconnect the power cord to the wall outlet and turn on the patient monitor. Leave the
patient monitor on until the battery is fully recharged and the orange battery charging
indicator LED turns off.
3.
Repeat the steps 1 and 2 once.
The battery is now conditioned and ready for use. However if either
Condition Battery A
or
Condition Battery B
message is still shown on the patient monitor screen, repeat the
conditioning cycle once more. If the problem persists, replace the battery.
NOTE: Refer to section
11.7. Error messages and codes
for more
detailed information about the battery status and operating condition.
10.4.6 Replacing a battery
Replace the battery in the following situations:
−
if either
Replace battery A
or
Replace battery B
message is displayed. This message
indicates that the full-charge capacity of the used battery has considerably degraded
compared to the design capacity of a new battery.
−
if either
Battery A failure
or
Battery B failure
message is displayed.
Remove the battery from the patient monitor and install a new battery according to the battery
installation instructions in the Hardware Installation section
NOTE: Refer to the section
and
11.7. Error messages and codes
for
more detailed information about the battery status and operating condition.
NOTE: Refer to the Troubleshooting chapter for more detailed information about the battery
diagnostics and error messages.
WARNING
Do not incinerate a battery or store at high temperatures. Serious injury or
death could result.
10.4.7 Battery recycling
This product contains Lithium-Ion batteries. At the end of their service life,
batteries in this product must be recycled or disposed in accordance with local
or national regulations. Do not dispose of batteries as trash or unsorted
municipal waste. Requirements and services for recycling of batteries vary
between countries.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......