![GE HEALTHCARE CARESCAPE Monitor B450 Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/ge-healthcare/carescape-monitor-b450/carescape-monitor-b450_technical-manual_571890071.webp)
Hardware installation
63
2062973-003
6.6 Connecting to the MC Network or the S/5 Network
WARNING
MISSED ALARMS - Do not use with iCentral software with Versions 5.0.3 and
earlier or with Mobile Care Server with Version 5.2 and earlier.
WARNING
INCORRECT CALCULATIONS - Using the System with the Aware Gateway
software version 1.4 or earlier could result in incorrect patient height and
weight information. This could lead to incorrect drug dose calculations,
hemodynamic calculations, or oxygenation calculations. Prior to installing
the System, please contact the GE Healthcare Aware Gateway HL7
Integration Engineering Team or your GE Healthcare service representative
to verify or update your Aware Gateway configuration.
WARNING
EXCESSIVE LEAKAGE CURRENT - Only devices that are specified compliant
with IEC 60950-1 or IEC 60601-1 may be connected to the Ethernet MC or IX
ports.
The patient monitor can be connected either to the wired MC Network or S/5 Network.
•
Connect the MC Network or the S/5 Network patch cable to the network connector
labelled as “MC” in the rear panel of the patient monitor.
6.7 Connecting to the IX Network
•
Connect the IX Network patch cable to the network connector labelled as “IX” in the rear
panel of the patient monitor.
6.8 Connecting to a Unity Network ID connectivity device
Ensure that the Unity Network ID connectivity device is properly mounted, installed and
configured prior to connecting it to the patient monitor.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......