Troubleshooting
161
2062973-003
On/standby button and alarm light issues
Alphanumeric Keyboard and barcode scanner issues
Speaker / audible alarm issues
Problem
Possible cause
Recommended action
On/Standby and the power
indicator LEDs are inoperative.
User interface cable is loose or
faulty.
Check user interface cable and replace it,
if necessary
User interface board is faulty.
Replace the user interface board.
Alarm light does not illuminate
when there is an alarm
condition on (audible alarms
work and alarm message is
visible)
Alarm light cable is loose.
Connect the alarm light cable to the DC/
DC board.
Alarm light board or alarm light
cable is faulty.
Replace the alarm light board
Alarm light does not illuminate
during power-on self test.
“Warm start”
It is less than 15 minutes since the last
power-up (warm start) and this normal
operation.
The alarm light illuminates during the
power-on self test only if it is a cold start
(more than 15 minutes from the previous
start-up).
Problem
Possible cause
Recommended action
Wrong character is displayed
when a key is pressed on
keyboard.
The keyboard locale is not
configured correctly.
Configure the keyboard locale correctly,
see
Wrong character is displayed
when a barcode is read.
The keyboard locale is not
configured correctly.
Configure the keyboard locale correctly,
see
The barcode reader’s language
configuration is incorrect.
Refer to the barcode reader manual.
Barcode reader does not read a
multi-field barcode correctly.
(i.e., the information is not
populated correctly to the fields
in the Admit menu).
Barcode reader parser
configuration is incorrect.
Configure the barcode settings, see
.
The barcode reader is
incompatible with the parser
configuration (field lengths, field
types, delimeters, symbologies
etc.).
Check the barcode information content
and compare it to the current parser
configuration.
Problem
Possible cause
Recommended action
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......