CARESCAPE Monitor B450
162
2062973-003
11.8.3
Incorrect system time
Audible alarms do not work.
Audible alarms are turned off
(See
Alarms Setup > Audible &
Visual.
)
Enable audible alarms.
Alarm volume is low.
Adjust alarm volume
(
Monitor Setup >
Sound Volumes
).
Speaker failure
Replace the speaker unit.
Speaker cable loose or faulty.
Check that the speaker cable is intact and
properly connected to the E-module
board.
Tone generator or audio
amplifier failure
Replace CPU board assembly.
Problem
Possible cause
Recommended action
System time is incorrect when
patient monitor is not
connected to network.
CPU timekeeper battery empty
Replace CPU timekeeper battery.
Time not configured properly
Configure date and time, see
System time is incorrect when
patient monitor is connected to
MC Network.
Network device time
synchronization error.
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device’s time. To
prevent potential time synchronization
issues, you should set the new device’s
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
iCentral time incorrect.
Configure iCentral time master.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......