CARESCAPE Monitor B450
114
2062973-003
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
Checkout procedure for wireless patient monitors
Check each wireless patient monitor according to the following procedure.
The purpose of this test is to ensure that each wireless patient monitor is correctly configured
1.
Disconnect the network cable from the MC Network connector, if connected.
2.
Check that the wireless network symbol and the signal strength symbol are shown
and that the signal strength is adequate.
3.
Select
Data & Pages – Other Patients
and select a patient from the list.
4.
Select
View
and verify that a window with parameters from another patient displays on
the left side of the screen. Check that the waveforms are continuous and there is no data
loss.
5.
Select
Close View
to close the window.
6.
Log in to Webmin.
7.
Select
Diagnostics > Ping
.
8.
Type the IP address of the patient monitor you just viewed and select
ping
. Check that the
patient monitor replies to the ping command, there are no lost packets and the maximum
latency is 250 ms.
NOTE: The patient monitor withstands a maximum packet loss of 5 packets per 1 million
and maximum latency of 250 ms without performance degradation.
9.
Select
to return to the main display.
10. Reconnect the MC Network cable back to the MC Network connector, if applicable.
Verification procedure for wireless MC Network infrastructure
Verify the wireless MC Network infrastructure according to the procedure described in
Appendix C.
The purpose of this test is to ensure the operation of the wireless network infrastructure
coverage area with a wireless CARESCAPE monitor.
8.3.11 IX printers
Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers > Print Test Page
.
3.
Select the IX Printer to test.
4.
Select
Save
.
5.
Verify that the test page was printed to the selected printer.
6.
Repeat steps 3 to 5 for all connected IX printers.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......