CARESCAPE Monitor B450
88
2062973-003
4.
In the
Remote Service Configuration
window, enter the applicable data:
5.
Select
Save
.
The changes will take effect immediately.
7.16.2 Enabling remote service agent/ connection
After the server has been configured for remote serviceability, the remote service agent must
be enabled for use.
1.
Log in to Webmin.
2.
Select
Configuration
>
Remote Service
.
3.
In the
Sub-Modules for Remote Service
menu, select
Control
.
4.
In the
Remote Service Control
window, enable or disable the
Service Agent
by
selecting
Enable
or
Disable
.
5.
Select
Save
.
The changes will take effect immediately.
HTTP Proxy Server Configuration
Item
Description
Comments
Address
- If this site uses an HTTP proxy server, a
specific site proxy server IP
Address
and
Port
number are required for the
Remote Service
communication to
work. Otherwise, select
None
.
- If the HTTP proxy server requires user
authorization, a specific
Username
, and
Password
is required. Otherwise, select
None
.
These values are
determined by the
customer.
Port
Username and
Password
Remote Service Configuration
Item
Description
Comments
System ID
Identifies the system to the GE back office
servers.
These values are read-only
and are unique.
Serial Number
Identifies the unit and is set at the time of
manufacture.
Enterprise URL
If required, designate the address of the GE
backoffice servers required to
communicate with the
Remote Service
Agent
.
This address should never
be changed unless explicit
instructions are given to do
so.
Enterprise Tunnel
URL
If required, designate the address of the GE
backoffice servers required to
communicate with the tunneling agent.
Protocol
Identifies the protocol used to
communicate with the enterprise servers.
This field is read-only and
cannot be changed.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......