Configuration
91
2062973-003
activation process with a delay during an active patient case, but the new settings will activate
only after the next case reset / patient discharged.
1.
Log in to Webmin.
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Activate
.
4.
In the
Activate Settings
window, select the settings that you want to activate:
•
Select
Cloned settings (clinical and platform)
to activate both the clinical and
platform settings.
•
Select
Cloned clinical settings
to activate only loaded clinical settings.
•
Select
Cloned platform settings
to activate only loaded platform settings.
NOTE: Activation of the cloned clinical or platform settings will not effect the following
settings in the target patient monitor:
−
Network
−
Unit and Bed Name
−
Host Asset Settings
−
Licenses
Remember to configure the unaffected settings manually in the target patient monitor.
NOTE: Check and manually modify the following WLAN settings after activating them in
the destination monitor:
−
RTS Threshold
−
Fragmentation Threshold
−
TXOP for Best Effort
−
TXOP for Background
•
Select
Factory default settings
to restore all platform and clinical settings back to
the factory defaults.
•
Select
US default clinical settings
to restore all clinical and platform settings back
to the US specific factory defaults.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......