Installation checkout
109
2062973-003
−
Polarity: REVERSED
−
Neutral: CLOSED
−
Earth (GND): OPEN
9.
Read and record the current leakage indicated on the safety tester.
10. Configure the safety analyzer as follows (SFC):
−
Polarity: REVERSED
−
Neutral: OPEN
−
Earth (GND): CLOSED
11. Read and record the current leakage indicated on the safety tester.
12. Configure the safety analyzer as follows (NC):
−
Polarity: REVERSED
−
Neutral: CLOSED
−
Earth (GND): CLOSED
13. Read and record the current leakage indicated on the safety tester.
14. Power off the device under test.
15. Repeat this test for all the connected parameter modules and patient connectors
Acceptance criteria in Normal Condition (NC):
•
All readings shall be less than or equal to 10
μ
A.
Acceptance criteria in Single Fault Condition (SFC) – earth open or one of the supply conductors
open at a time:
•
All readings shall be less than or equal to 50
μ
A.
8.2.9 Patient (sink) leakage current tests
This procedure measures the leakage current from an applied part connector of the device to
ground. Perform the test in Normal Condition (NC) with normal and reverse polarity.
NOTE: Refer to the instructions delivered with the safety analyzer to perform this test.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......