CARESCAPE Monitor B450
68
2062973-003
3.
Enter the new hostname in the
Change Value to
column.
NOTE: The hostname is a unique, 4 to 32 character long identifier of a patient monitor in
the network. Use alphanumeric characters A-Z, a-z, 0-9. The hostname may include also
characters “-” and “_”, but it can't start or end with these characters.
4.
Select
Save
.
The hostname configuration will take effect immediately.
7.2.2 Selecting and configuring CARESCAPE Network
To select CARESCAPE Network MC for the real-time network infrastructure:
1.
Log in to Webmin.
2.
Select
Configuration >
Network > Wired Interfaces
.
The current network configuration is shown in the
Present Configuration
table.
3.
Below
Select Network Type
in the
Network Configuration
window, select the applicable
network type:
CARESCAPE Network
.
4.
Select
Next
to proceed on network configuration.
NOTE: When
CARESCAPE Network
is selected, the S/5 Network is disabled.
The patient monitor may be configured to the MC Network, or IX Network, or both.
NOTE: The following IP addresses are reserved and not valid: 172.16.254.254, 172.16.255.255,
172.18.254.254, 172.18.255.255, 192.168.250.x, 192.168.252.x, 192.168.253.x, 192.168.254.1
and 192.168.254.2.
5.
In the
Network Configuration
window below
MC Network.
a.
Enter a
Static IP
address.
b.
Enter a valid
Netmask
level.
c.
Select the applicable
Speed & Duplex
option.
6.
If separate MC Network segments are connected together by a router, configure the
router information below the MC static route:
a.
Enter
Destination Address
.
b.
Enter
Destination Netmask
.
c.
Enter an
MC Gateway.
7.
Below
IX Network,
select
DHCP
or
Manual Configuration
.
If
Manual Configuration
is selected, enter the following information in the entry fields:
a.
Enter a
Static IP
address.
b.
Enter a valid
Netmask
level.
c.
Enter a valid
Default Gateway
.
d.
Enter valid
DNS Server
addresses if applicable.
8.
Select the applicable
Speed & Duplex
option.
9.
Select
Save
.
Network connection(s)
Steps to be performed
MC Network only
5, (6) and 9
IX Network only
7, 8 and 9
Both MC and IX Network
5, (6), 7, 8 and 9
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......