CARESCAPE Monitor B450
62
2062973-003
patient, the operator, or the environment as a result. In those instances
where there is any element of doubt concerning the safety of the connected
devices, the user must contact the manufacturers concerned (or other
informed experts) for proper use. In all cases, safe and proper operation
should be verified with the applicable manufacturer's instructions for use,
and system standards IEC60601-1-1 must be complied with.
WARNING
Do not under any circumstances remove the grounding conductor from the
power plug. Always check that power cord and plug are intact and
undamaged.
WARNING
POWER SUPPLY — The device must be connected to a properly installed
power outlet with protective earth contacts only. If the integrity of the
protective earth conductor is in doubt, disconnect the monitor from the
power line, and use it with the battery power. All devices of a system must
be connected to the same power supply circuit. Devices which are not
connected to the same circuit must be electrically isolated when operated.
CAUTION
POWER REQUIREMENTS - Before connecting the device to the power line, check
that the voltage and frequency ratings of the power line are the same as those
indicated on the device's label. If this is not the case, do not connect the system
to the power line until you adjust the device to match the power source. In
U.S.A., if the installation of this equipment will use 240V rather than 120V, the
source must be a center-tapped, 240V, single-phase circuit. This equipment is
suitable for connection to public mains as defined in CISPR 11.
If a non-medical grade display is used as a display within the patient environment it must
always be powered from an additional transformer providing at least basic isolation.
1.
Connect power cords to the mains power supply inlet and to a wall outlet on all system
components that require AC mains power input.
2.
Do not power on any devices.
NOTE: Be sure that all power cords are securely connected and that they are routed through
the retaining clips or cable clamps, as applicable.
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......