CARESCAPE Monitor B450
174
2062973-003
more than three times. Use only new screws attaching into light metal parts. Take
advantage of existing thread pattern cut by turning the screw counterclockwise until it
drops into the existing thread pattern.
•
We have added the maximum recommended torque value to be used for each screw and
nut in reassembly. These values are listed in brackets after each disassembly step.
•
When attaching self-tapping screws to light metal parts without existing threads (new
light metal FRU parts), you should use a higher torque than is recommended for
reassembled parts, but still not more than 1.6 Nm.
NOTE: It is not required to a use torque wrench or torque screwdriver when reattaching the
patient monitor using the recommended standard hand tools. Just ensure visually that the
screws are properly attached, but avoid over tightening by using excess force as this may
damage the existing thread patterns. However, when using battery-operated tools, ensure that
it is equipped with torque limiter and the torque is properly adjusted.
12.1.3
Required tools
NOTE: Use torque wrench and torque screwdriver to comply with the given torques.
WARNING
Due to possible high voltage present, use an insulated screwdriver at all
times.
12.1.4
Before disassembly
WARNING
PATIENT MONITORING INTERRUPTION — Make sure a patient is not being
monitored while servicing the equipment.
WARNING
DISCONNECTION FROM MAINS - When disconnecting the device from the
power line, remove the plug from the wall outlet first. Then you may
disconnect the power cord from the device. If you do not observe this
sequence, there is a risk of coming into contact with line voltage by
inserting metal objects, such as the pins of leadwires, into the sockets of the
power cord by mistake.
WARNING
SAFETY GROUND - Remove power cord from the mains source by grasping
the plug. Do not pull on the cable.
- insulated Torx T6 and T10 screwdrivers
- insulated flat blade screwdriver (width 2.5 mm
/ 0.1 in)
- pliers
- antistatic ESD wristband
Содержание CARESCAPE Monitor B450
Страница 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Страница 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Страница 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Страница 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Страница 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Страница 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Страница 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Страница 127: ...Theory of operation 119 2062973 003 9 2 DC DC Board Figure 4 CARESCAPE Monitor B450 DC DC Board Block Diagram ...
Страница 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Страница 175: ...Troubleshooting 167 2062973 003 ...
Страница 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Страница 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Страница 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Страница 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Страница 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Страница 235: ......