![Amsco Century Скачать руководство пользователя страница 95](http://html1.mh-extra.com/html/amsco/century/century_operators-manual_2936190095.webp)
9-7
Programming Operating Parameters
Operating Instructions
129373-496
•
Pressing
DOOR SEAL
advances display to screen #54. Enter door seal
phase value using the numeric keypad. Value appears on display as it is
being entered.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#53.
54
CHANGE VALUES - TOO LONG IN DOOR SEAL
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN
DOOR SEAL
30:00
(Minutes:Seconds)
The too long in
door seal is the
allowable amount of
time that the door
has to seal before an
alarm occurs.
•
Pressing
DOOR UNSEAL
advances display to screen #55. Enter door
unseal phase value using the numeric keypad. Value appears on display
as it is being entered.
55
CHANGE VALUES - TOO LONG IN DOOR UNSEAL
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN
DOOR UNSEAL
30:00
(Minutes:Seconds)
The too long in
door unseal is the
allowable amount of
time that the door
has to unseal before
an alarm occurs.
•
Pressing
JACKET CHARGE
advances display to screen #59. Enter jacket
charge phase value using the numeric keypad. Value appears on display
as it is being entered.
59
CHANGE VALUES - TOO LONG IN JACKET CHARGE
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN
JACKET CHARGE
30:00
(Minutes:Seconds)
The too long in
jacket charge is the
allowable amount of
time that the jacket
has to charge to the
set point before an
alarm occurs.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#53.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#53.