Operator Manual
Troubleshooting
Occurs if liquid cycle
fast exhaust rate is too
fast.
12.2.9 Exhaust Rate Too
Fast
Screen with Service Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
241
EXHAUST RATE TOO FAST
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
TRY TO EXHAUST CHAMBER ACCORDING
TO OPTIMAL COOLING RATE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
242
EXHAUST RATE TOO FAST
→
→
→
→
→
CHAMBER EXHAUSTED FASTER
THAN THE EXPECTED RATE.
CAUSES AND CORRECTION:
1.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S40
2.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S03
3.
CHAMBER STEAM TRAP MALFUNCTION
→
→
→
→
→
REPAIR
4.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
CLEAR
EXIT
FEED
ALARM
Chamber Pressure/
Temperature Failure
(Continued)
Alarm
Description
Screen with Service Instructions
Screen with Operator Instructions
STATUS..
SERVICE INFO: CHAMBER
220
PRESSURE/TEMPERATURE FAILURE
→
→
→
→
→
PRESSURE OR TEMPERATURE OUTSIDE
NORMAL STEAM RANGE
CAUSES AND CORRECTION:
1.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
2.
TRANSDUCER, CP, MALFUNCTION
→
→
→
→
→
REPAIR
3.
RTD PROBE, RTD1, MALFUNCTION
→
→
→
→
→
REPAIR
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD POWER STATUS
LEDs
→
→
→
→
→
REPLACE CONTROL BOARD
→
→
→
→
→
RECALIBRATE
PAPER
SERVICE
EXIT
FEED
MODE