12-18
129373-496
Operator Manual
Troubleshooting
Screen with Operator Instructions
Occurs if door seal
pressure does not
drop below 5 psig
(0.34 Pbar) within
allotted time.
12.4.4 Too Long To
Unseal Door
Screen with Service Instructions
Occurs if chamber
pressure reading is
outside the normal
range.
12.4.5 Chamber
Pressure
Transducer
Failure
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
229
TOO LONG TO UNSEAL DOOR
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS UNSEALED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
3. If load must be removed, refer to
emergency door operation procedure
in operating manual.
Alarm
Description
Screen with Operator Instructions
STATUS..
SERVICE INFORMATION:
230
TOO LONG TO UNSEAL DOOR
→
→
→
→
→
DOOR SEAL PRESSURE NOT BELOW 5 psig
WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
→
→
→
→
READJUST PS1(PS2)
→
→
→
→
→
REPAIR PS1(PS2)
2.
SEAL NOT RETRACTING
→
→
→
→
→
CHECK SEAL STEAM
→
→
→
→
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S37(S38)
→
→
→
→
→
REPAIR S35(S36)
PAPER
SERVICE
EXIT
FEED
MODE
STATUS
PAPER
SILENCE
SERVICE
ALARM
FEED
HELP
STATUS..
ALARM! CHAMBER
217
PRESSURE TRANSDUCER FAILURE
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE